4 Operation
4.4 Dentist element
Dentsply Sirona
Operating Instructions Simulation Unit Sim Multi Bench
62
66 82 384 D3684
D3684.201.01.02.02 07.2019
Purge
1.
Press the
Endo/ Purge key.
Ä
The endodontics intensity (Endo mode) is displayed instead of
ultrasonic intensity (ultrasonic mode).
1
2
Setup
3
2.
Use favorite keys 1 and 3 to set the desired endodontics intensity
(En 1 to En 5).
Pressing the
Endo / Purge key switches back to ultrasonic mode.
The intensity settings cannot be saved on the favorite keys in Endo
mode.
4.4.10
Cavitron scaler
The simulation unit can be equipped or retrofitted with the built-in scaler
from Cavitron.
Installation is done by the service technician. Contact your dental depot.
The Cavitron built-in scaler is used to remove plaque.
IMPORTANT
Please also read the operating instructions for the "Cavitron Built-In
Ultrasonic Scaler, Model G139 with Cavitron Steri-Mate 360°
Handpiece“. They are enclosed with the retrofit kit.
4.4.10.1
Safety Instructions
To prevent injuries, a tip protector is included with the handpiece.
CAUTION
Ultrasonic tips are are very sharp.
There is a risk of injuring the hand when the scaler is in the holder.
➢ Always attach the tip protector to the scaler for protection as soon
as you return the handpiece to the holder.
CAUTION
Ultrasonic tips from other manufacturers do not guarantee safe
operation.
➢ Use only ultrasonic tips from Dentsply Sirona.
CAUTION
The Cavitron scaler should not be used if the incoming treatment
water is warmer than 25 °C / 77 °F.
The spray water in the handpiece can heat up further due to
ultrasonic vibrations.
➢ Check the temperature of the water.