![Dentsply Sirona Sim Multi Bench Скачать руководство пользователя страница 12](http://html1.mh-extra.com/html/dentsply-sirona/sim-multi-bench/sim-multi-bench_operating-instructions-manual_2483697012.webp)
2 Safety instructions
2.9 Vacuum system
Dentsply Sirona
Operating Instructions Simulation Unit Sim Multi Bench
12
66 82 384 D3684
D3684.201.01.02.02 07.2019
2.9
Vacuum system
The suction removal of aluminum and other metal oxides from blasting
devices via the automatic separator built into the simulation unit is
prohibited! This would result in extreme wear and clogging of the
vacuum and drain lines.
A separate vacuum system must be used in connection with metal
oxide blasting devices. Simulation units equipped with a central
standard wet suction are generally suitable for suction removal of the
above material. However, make sure to observe the instructions
provided by the manufacturer of your vacuum system.
No restrictions apply when using salt blasting devices in connection with
Sirona simulation units. However, in such cases, make sure that the
system is subsequently flushed with an adequate amount of water.
To clean the vacuum system, a suction hose connection is optionally
integrated in the supply module. Water is pumped into a container
behind the suction hose intake and suctioned off from there, see
“Vacuum system”.
The unit can be optionally operated with an air jet pump (Air Venturi)
instead of a suction device.
2.10
Intermittent operation
The motors of the treatment instruments are designed for intermittent
operation corresponding to the dental mode of treatment.
IMPORTANT
The procedures required for the following treatment instruments are
described in the separate pertaining operating instructions. The
instructions as well as all warning and safety information they contain
must also be followed.
2.11
Ventilation slots
Under no circumstances may the ventilation slots on the unit be
covered, otherwise the air circulation will be obstructed. This can cause
the unit to overheat.
Do not spray liquids such as disinfectants into the ventilation slots or
openings in the supply module or dentist element. This may lead to
malfunctions. Use only wipe disinfection in these areas.
Содержание Sim Multi Bench
Страница 1: ...New as of 07 2019 Simulation Unit Sim Multi Bench Operating Instructions USA Canada English ...
Страница 99: ......