4 Operation
4.4 Dentist element
Dentsply Sirona
Operating Instructions Simulation Unit Sim Multi Bench
38
66 82 384 D3684
D3684.201.01.02.02 07.2019
4.4.2
Fixed keys and displays
Operating light
Switches the operating light on, to the composite function, or off.
The composite function is used to avoid premature curing of the
composite materials.
In the LEDview Plus, the setting context or dialog appears when the key
is pressed (> 2 s).
The brightness of the light and the distance from which the non-contact
sensor should react to movements can be adjusted.
With the LEDlight Plus, the brightness is programmed using the
contactless sensor.
In the LEDview Plus LEDview Plus, the color temperature of the light
can also be adjusted.
For details, please refer to the section "Operating light [→ 66]".
Counterclockwise rotation / User profile
Setting the direction of rotation
A/B
The direction of rotation can be changed only with the motor stopped.
ü
For a removed motor: switching the counterclockwise rotation on/
off.
➢ Press the Counterclockwise rotation/user profile fixed key.
Ä
The Counterclockwise rotation status display lights up when
counterclockwise rotation is set.
Selecting a user profile
A/B
If the status display user profile lights up, user profile B is pre-selected;
if it does not light up, user profile A. The user profile used last is
automatically loaded when the simulation unit is switched on.
ü
With instruments in place: changing the user profile.
➢ Press the Counterclockwise rotation / User profile key.
Ä
The status display user profile lights up or goes off.
This toggles between the user profiles.
ü
With motor removed and torque control switched on:
➢ Press the button longer (> 2 s),
Ä
Switches AutoReverse function ON/OFF when the maximum
torque is reached.
➢ Press the format button briefly (< 2 s)
Ä
Switches counterclockwise rotation ON/OFF
Depending on the context, an ongoing process can be stopped and
switched to the standard operating context, e.g. ending the sanitizing
process.
Содержание Sim Multi Bench
Страница 1: ...New as of 07 2019 Simulation Unit Sim Multi Bench Operating Instructions USA Canada English ...
Страница 99: ......