4 Operation
Dentsply Sirona
4.5 Accessories for the operating light
Operating Instructions LEDview Plus
Mounting the camera holder
Fasten the camera to the camera holder with the tripod screw before
mounting the camera holder onto the operating light. Make sure that the
camera is aligned correctly and is securely positioned.
✔
The accessory interface is installed on the operating light.
✔
The operating light's joint brakes are pre-set to the individual weight
of the camera.
✔
The camera is fastened to the camera holder with the tripod screw.
CAUTION
Danger of electric shock
If the cable for the camera needs to be laid along the support arm
system then laying the cable inside the support arm panel can result in
the insulation wearing out and therefore in an electric shock.
➢ Always lay the cable outside of the panel for the support arm
system.
IMPORTANT
In order to ensure that the support arms can be disinfected, Dentsply
Sirona recommends using a camera with an internal power supply
(rechargeable storage battery) and an internal memory option (memory
card). This means that no additional cabling is required externally on the
support arm system.
IMPORTANT
If the operating light's joint brake is not set or is not set correctly, the
operating light may be subject to a tilting movement when the camera is
mounted.
Have the operating light's joint brake adjusted by a service engineer to
the individual weight of the camera used by you.