![Dentsply Sirona CEREC Omnicam AC Скачать руководство пользователя страница 92](http://html.mh-extra.com/html/dentsply-sirona/cerec-omnicam-ac/cerec-omnicam-ac_service-manual_2483632092.webp)
4 Repair
4.3 Replacing parts (without removing covers)
Dentsply Sirona
Service Manual CEREC Primescan AC, Primescan AC, CEREC Omnicam AC, Omnicam AC
92
66 81 832 D3696
D3696.076.01.02.02 05.2019
4.3.2.3.2
Removing and inserting the control unit cable
ü
The monitor is removed; see "Replacing the monitor [→ 85]".
1.
Raise the complete control unit carefully upwards slightly (see
"Removing and inserting the control unit [→ 89]").
2.
Disconnect the plug connections of the control unit cable (A) and
remove the control unit cable.
NOTE
When performing the following installation steps, make sure that all
plug connections snap in place correctly.
3.
Route the new control unit cable (A) and re-establish the plug
connections.
4.
Attach the control unit and screw it down (see "Removing and
inserting the control unit [→ 89]").
5.