4 Repair
4.3 Replacing parts (without removing covers)
Dentsply Sirona
Service Manual CEREC Primescan AC, Primescan AC, CEREC Omnicam AC, Omnicam AC
72
66 81 832 D3696
D3696.076.01.02.02 05.2019
1.
Slide the O-ring at one point into the O-ring groove.
2.
Starting from this point, carefully use your thumb and index finger to
push the O-ring into the groove (in a similar way to pushing it out
but without any force).
3.
Guide your thumb and index finger in such a way along the groove
until the O-ring is fully positioned in the groove.
NOTE
The O-ring can be damaged
Ensure that the O-ring fits in the groove without twisting in order to
prevent damage to the O-ring when fitting the sleeve.
IMPORTANT
Replacement O-rings
Replacement O-rings are available with the scanner or can be
ordered under REF 66 80 974.