4 Repair
4.2 Removing covers and holders
Dentsply Sirona
Service Manual CEREC Primescan AC, Primescan AC, CEREC Omnicam AC, Omnicam AC
54
66 81 832 D3696
D3696.076.01.02.02 05.2019
4.2.1.5
Right cover
1.
Unscrew the 3 screws (A, TX 20).
2.
Flip the footswitch down.
3.
Unscrew the right screw (B).