Dentsply Sirona
Service Manual CEREC Primescan AC, Primescan AC, CEREC Omnicam AC, Omnicam AC
4 Repair
4.3 Replacing parts (without removing covers)
66 81 832 D3696
D3696.076.01.02.02 05.2019
73
4.3.2
Replacing parts (by removing the service cover)
4.3.2.1
Replace the monitor kinematics
Use the parts available in the "6720135 Kinematik ERS SET" spare
parts to replace the monitor kinematics.
4.3.2.1.1
Removing the covers
1.
Remove the monitor (see "Replacing the monitor [→ 85]") and
place it on a soft and clean base as shown.
2.
Unscrew 2 screws (A, M3 X 6 Precote, TX 10).