5 Settings and adjustments
5.1 Set the monitor kinematics
Dentsply Sirona
Service Manual CEREC Primescan AC, Primescan AC, CEREC Omnicam AC, Omnicam AC
120
66 81 832 D3696
D3696.076.01.02.02 05.2019
2.
Unscrew the 2 M12 nuts (A) and screw them away from the bearing
pedestal.
3.
Apply Loctite carefully on the gear shaft between the M12 nuts and
bearing pedestals.
4.
Screw the 2 M12 nuts back onto the bearing pedestal.
5.
Loosen the 2 screws (B, M4 X 35). Leave the other ones screwed in
so that the centering does not change!
6.
Retighten one M12 nut (A) in a clockwise direction (move wrench
up). Start with a 1/2 rotation.
Ä
The disk springs are pressed together -> the tightness
increases.