Dentsply Sirona
Service Manual CEREC Primescan AC, Primescan AC, CEREC Omnicam AC, Omnicam AC
3 Scanner
3.3 Scanner test and diagnostics
66 81 832 D3696
D3696.076.01.02.02 05.2019
41
3.3.1.1.2
Possible symptoms
Below you will find a list with possible symptoms and a sequence of
solutions which are recommended in each respective case. All
symptoms refer to the CEREC Primescan / Primescan scanner.
#
Symptom
Solution sequence
S1
Scanner is not recognized by the user software.
L1: Restart user software and/or PC.
L3: Disconnect / plug in scanner cable
L8: Identify scanner as USB device.
L4: Start analysis with DrPrimescan.
L5: Test available licenses.
Optical impression with scanner is not possible in
the user software.
See S1
S2
Scanner is not recognized by DrPrimescan.
L2: Disconnect PC from the power supply.
L1: Restart user software and/or PC.
L3: Disconnect / plug in scanner cable
L8: Identify scanner as USB device.
L19: Check (if possible) whether the supply board is
working.
L11: Replace supply board.
L15: Replace the scanner.
S3
2D image quality inadequate:
● Tint
● Image too dark / too bright
L20: Setting the right exposure mode
(extraoral/intraoral).
L6: 3D calibration of the scanner
L7: Color calibration of the scanner
S4
Color impression of the 3D model inadequate:
● Tint
● Image too dark / too bright
See S3
S5
The scanner is not recognized as a USB device
with the name "MHT Optic Research AG – 3D scan-
ner".
L2: Disconnect PC from the power supply.
L1: Restart user software and/or PC.
L3: Disconnect / plug in scanner cable
L8: Identify scanner again as USB device.
L9: Restore process
S6
No 3D Preview available. Live image is OK.
L1: Restart user software and/or PC.
L6: 3D calibration of the scanner
L16: Check / install graphic card driver.
L9: Restore process
S7
3D Preview stops, acquisition process cannot be re-
sumed.
Indicator: Audio feedback.
Live image is OK.
Focus for resumption of last good exposure (e.g.
measured occlusal surface)
L1: Restart user software and/or PC.
L20: Setting the right exposure mode
(extraoral/intraoral).
L6: 3D calibration of the scanner
L16: Check / install graphic card driver.
L9: Restore process