Dentsply Sirona
Service Manual CEREC Primescan AC, Primescan AC, CEREC Omnicam AC, Omnicam AC
4 Repair
4.3 Replacing parts (without removing covers)
66 81 832 D3696
D3696.076.01.02.02 05.2019
91
2.
Attach the USB connector and the connector of cable L1 and route
the cable as shown.
3.
Bring the locking plates into the correct position for the insertion of
the control unit. The gap (A) must be open (as shown).
4.
Install the control unit. If the control unit gets stuck or is difficult to
move once inserted into the frame, correct the position of the plates
screwed onto the control unit.
5.
Push the left locking plate forwards up to the stop and then screw it
back on again with the screw. Perhaps help things along by
simultaneously pressing the control unit downwards slightly on the
relevant side.
6.
Push the right locking plate forwards up to the stop and then screw
it back on again with the screw.
7.
Reinstall the monitor; see "Replacing the monitor [→ 85]".