Motor Handpiece Error Code
66/74
F1902142.EN / 05 / 2015 / updated 02/2016
The Motor
START/STOP
button sensor
is not working
properly
Bluetooth radio
module failure
Calibration
failed
Lubricate the contra-angle (see
Lubrication of the Contra-Angle
) and
perform the calibration again.
If the error persists, try to replace the contra-
angle and perform the calibration again.
Battery charge
not sufficient to
perform a
firmware
upgrade.
Charge the device until the BATTERY
indicator turns green. Restart the firmware
update.
General
firmware failure
State
Failure
Solution