Cleaning, Disinfection and Sterilization
F1902142.EN / 05 / 2015 / updated 02/2016
57/74
7.4.2
Manual Reprocessing
If you clean and disinfect the contra-angle manually, heed the
instructions below and proceed as follows:
N°
Operation
Remarks
A
Manual cleaning
For 5 minutes, rinse and
brush under running
deionized water (DI), or
water that has this
degree of purity (<38°C
(100.4°F)).
Remove any liquid
residues (ultra-absorbent
cloth, particle-free
compressed air).
Use a cleaning tool (brush).
B
Manual disinfection
Thoroughly disinfect
surfaces with a soaked
wipe, making sure the
entire surface is wet, and
allow to act (exposure
time > 30 minutes).
For disinfection purposes, Dentsply Maillefer
recommends the use of mikrozid
®
AF wipes
manufactured by Schülke.
If large surfaces have to be treated, use more
than one wipe where necessary.
Make sure the surface is completely dry after disinfection.
Only use surface disinfectants that are certified by officially
recognized institutes, do not contain chlorine and have been declared
aldehyde-free.
Please refer to the manufacturer’s information on use of the
disinfectant.