Cleaning, Disinfection and Sterilization
62/74
F1902142.EN / 05 / 2015 / updated 02/2016
7.4.5.1
Approved Sterilization Procedure
N°
Operation
Remarks
A
Sterilization
Steam sterilization with
fractionated pre-vacuum
method B.
Sterilization at 132°C
(269.6°F) for at least 4
min. or at 121°C
(249.8°F) for at least
30 min.
Comply with your national guidelines, standards
and requirements for cleaning, disinfection and
sterilization.
Steam sterilization in accordance with EN 13060
or EN 285.
The fast sterilization procedure and the sterilization procedure with
non-packaged accessories are not permissible. In addition, do not
use hot-air sterilization, radiation sterilization or sterilization with
formaldehyde, ethylene oxide or plasma.
B
Storage
Store sterile contra-
angles in a dry, dust-free
place at room
temperature.
Before resuming operation: wait until the contra-
angle attachment has cooled down and is
completely dry.
Contra-angles must be checked for damage before they are reused.
Damaged products must be replaced.