Cleaning, Disinfection and Sterilization
58/74
F1902142.EN / 05 / 2015 / updated 02/2016
7.4.3
Machine-based Reprocessing
If you clean and disinfect the contra-angle in a machine, heed the
instructions below and proceed as follows:
N°
Operation
Remarks
A
Automatic cleaning /
disinfection
Cleaning at 65°C (149°F)
holding time 5 min.
Disinfection at 90°C
(194°F) holding time
5 min.
Use washer-disinfector conforming to
EN ISO 15883, e.g. Belimed WD, programme:
neutral.
Use cleaning solution: e.g. 0.5% (v/v) deconex
24 LIQ, Borer Chemie.
Use neutralizing solution: e.g. 0.25% (v/v)
deconex 26 plus, Borer Chemie.
Deionized water (DI) or water that has this
degree of purity is recommended for cleaning.
Make sure the contra-angle attachment is completely dry after
thermal disinfection.
Remove any residual liquid with particle-free compressed air.
Drying is performed automatically during the drying cycle in the
washer-disinfector.