Troubleshooting
68/74
F1902142.EN / 05 / 2015 / updated 02/2016
Lost connection
between the
handpiece and the
iPad
®
.
iPad
®
switched off (empty battery).
The handpiece keeps the last used settings for 5 min. You
can either recharge the iPad
®
and reconnect the motor
handpiece or use it in stand-alone mode.
iPad
®
Bluetooth
®
switched off.
Switch on the Bluetooth
®
on the iPad
®
and reconnect the
6.5.4 Activating the Bluetooth
6.7.3 Connecting the Motor Handpiece
).
Motor handpiece far from the iPad
®
.
Place the motor handpiece near the iPad
®
and re-connect it
6.7.3 Connecting the Motor Handpiece
).
The contra-angle
cannot be calibrated.
The calibration procedure may have been interrupted by
increased resistance in the contra-angle.
1. If calibration has been interrupted, calibrate the motor
handpiece again to rule out the possibility of a motor fault.
2. Clean and lubricate the contra-angle (see
Disinfection and Sterilization of the Contra-Angle
3. Start the calibration procedure again.
Motor handpiece is
getting hot.
Overheating.
The device switches off automatically if it overheats. Allow
the device to cool down and start again the motor handpiece.
A reciprocating
instrument has
become jammed in
the root canal.
Treatment was not performed properly.
Remove the reciprocating instrument with pliers by rotating
clockwise carefully.
A rotary instrument
has become jammed
in the root canal.
Treatment was not performed properly.
Select reverse rotation, start the motor handpiece and rotate
the instrument out of the root canal carefully.
Problem
Cause / Solution