Cleaning, Disinfection and Sterilization
52/74
F1902142.EN / 05 / 2015 / updated 02/2016
7
Cleaning, Disinfection and Sterilization
7.1
Foreword
Regular maintenance of
X-SMART IQ™
is essential to ensure that work
performed is completely hygienic and that the device will have a long
service life.
The components of
X-SMART IQ™
have not been disinfected or
sterilized when they are delivered. Please perform the necessary steps
before the first treatment.
You are responsible for the sterility of accessories; always ensure that
only validated methods of cleaning/disinfection and sterilization are
used, that regular maintenance and inspections are performed on the
equipment (disinfector, sterilizer) and that the validated parameters are
maintained in every cycle.
Comply with your national guidelines, standards and requirements for
cleaning, disinfection and sterilization.
7.2
Single Use
7.3
Disinfection
must be cleaned and subjected to a
wipe-down disinfection before the first treatment and immediately after
every subsequent treatment. They are not approved for machine-based
cleaning (washer-disinfector) and sterilization.
Wear protective clothing during all cleaning, disinfection and
sterilization processes.
The use of the handpiece sleeve is mandatory. This product is
designed for single use and has to be disposed of and changed after
every treatment. In the event of non-compliance, the contamination
risk increases.