Cleaning, Disinfection and Sterilization
F1902142.EN / 05 / 2015 / updated 02/2016
59/74
7.4.4
Lubrication of the Contra-Angle
It is essential to lubricate the contra-angle after every internal cleaning
procedure, before every sterilization or after 30 min. of use, but at least
once a day.
Please perform the following steps to lubricate the contra-angle:
When you lubricate the contra-angle take care to ensure that no oil
gets into the motor handpiece. Contamination of the motor handpiece
with oil can damage it and have a negative impact on safety when it is
being operated.
Lubricate the contra-angle immediately after thermal disinfection.
Use the spray nozzle included to lubricate the contra-angle.
Dentsply Maillefer recommends reprocessing the contra-angle using
a fully automatic reprocessing device, e.g. the Assistina from W&H.
N°
Action
A
Remove the spray nozzle cap.
B
Attach the spay nozzle to the oil can (e.g. W&H Service Oil MD-400).
C
Lubricate the contra-angle according to the information in the Instructions for
Use of the oil spray can.
A
B
C