39
Confirm with ENTER. The following screen will appear:
PROBE CALIBRATION
CO CALIBRATION
1) Cal zero
2) Sensitivity set
▼▲
select
<ESC> exit/cancel
<ENTER> confirm
Confirm with ENTER. The following screen will appear:
PROBE CALIBRATION
CO CALIBRATION
CAL ZERO
Cal zero 0 ppm
<ESC> exit/cancel
<ENTER> confirm
•
Wait about 15 minutes before continuing.
•
Supply the gas adjusting the bottle flow meter to get a constant flow from 0.1 to 0.2 l/min.
•
Press ENTER and wait the two minutes necessary for calibration without changing the
working conditions.
•
At the end, close the bottle spigot and remove the cap from the CO sensor.
•
Insert the protection grid.
Replacement of the CO sensor:
In normal usage conditions, the CO sensor has an average life longer than 5 years. If you need to
replace the CO sensor, order a new sensor (code
ECO-SURE-2E CO
).
To replace the CO sensor, follow this procedure:
1.
Switch the instrument off.
2.
On the P37AB147 probe, open the
sensor-holder small door using a
screwdriver and extract the
exhausted CO sensor.
3.
Take a new CO sensor and note
down the number printed on the
edge that indicates its nA/ppm
sensitivity.
4.
Insert the new sensor electrodes in
the terminals.
5.
Turn the instrument on, and wait
for at least 5 minutes so as to
stabilize the measurement.