24
4.3.1 The Baud Rate
The
Baud Rate
indicates the speed used for the serial communication with the PC.
To set the
Baud Rate
, proceed as follows:
1.
Use the arrows
▲▼
to select the item;
2.
Press
ENTER
: You will get the following message:
COMMUNICATION MENU
SERIAL
SET BAUDRATE
Baud Rate: 38,400
▲▼
set
<ESC> exit
<ENTER> confirm
3.
Use the arrows
▼▲
to set the value;
4.
Press
ENTER
to confirm and return to the previous page, or press
ESC
to cancel the change and
exit the menu item;
5.
Press
ESC
over and over to exit from the submenus;
6.
Press
MENU
to exit immediately from the main menu.
WARNING: The communication between instrument and PC (or serial port printer) only
works if the instrument and PC baud rates are the same
. If the USB connection is used, this
parameter value is automatically set.
NOTE: When setting the baud-rate, check the printer speed.
4.3.2 Print Interval
To set the
Print Interval
, proceed as follows:
1.
Use the arrows
▲▼
to select the item;
2.
Press
ENTER
: You will get the following message:
COMMUNICATION MENU
SERIAL
PRINT INTERVAL
h:mm:ss (1h max)
0:00:00
▼▲
set
<ESC> exit
<ENTER> confirm
3.
Use the arrows
▼▲
to set the value;
4.
Press
ENTER
to confirm and return to the previous page, or press
ESC
to cancel the change and
exit the menu item;
5.
Press
ESC
over and over to exit from the submenus;
6.
Press
MENU
to exit immediately from the main menu.
The print interval can be set from 0 seconds to one hour: