ROBE CALIBRATION
The instruments and probes are calibrated in the factory; no calibration is usually required by the
user. However, you can perform a new calibration.
With the HD37AB1347 instrument, you can perform the calibration of the following connected
probes:
•
With the
P37AB147
and
P37B147
SICRAM probes connected to input 1 Indoor Air
Quality, you can perform the calibration of the RH (Relative Humidity), CO (Carbon
Monoxide – only
P37AB147
), and CO
2
(Carbon Dioxide) sensors.
•
With the
temperature
and
humidity
combined SICRAM probes connected to input 1
Indoor Air Quality, you can perform the calibration of the RH (Relative Humidity) sensor.
No calibration is allowed for the Temperature sensor, and for the Hot-Wire and Vane sensors.
To calibrate the probes correctly, a knowledge of and abiding by the physical phenomena on which
the measurement is based is fundamental: this is the reason why it is recommended to abide by what
is reported below carefully and only to perform new calibrations if technically proficient.
PROBE CALIBRATION
1) Calibration CO
2
2) Calibration CO
3) Calibration RH
▼▲
select
<ESC> exit/cancel
<ENTER> confirm