background image

 

HD35 

-   84   -

 

V1.12

 

9.10

 

TECHNICAL CHARACTERISTICS OF DATA LOGGERS IN WATERPROOF HOUSING

 

Transmission frequency 

868 MHz, 902-928 MHz or 915.9-929.7 MHz according to the model 

Antenna 

Internal (default). On request, external fixed or with 3 m cable. 

Transmission range 

In open field: 

300 m (E, J)/ 180 m (U) with internal antenna towards base unit (except 

HD35APD) and repeaters. 

180 m (E, U) with internal antenna towards base unit HD35APD. 

> 500 m (E, J, U) with external antenna towards base unit (except 

HD35APD) and repeaters. 

180 m (E, U) with external antenna towards base unit HD35APD. 

(could be reduced in presence of obstacles or adverse atmospheric 

conditions

Measuring interval 

(*)

 

1, 2, 5, 10, 15, 30 s / 1, 2, 5, 10, 15, 30, 60 min 

Logging and 

transmission interval 

(*)

 

1, 2, 5, 10, 15, 30 s / 1, 2, 5, 10, 15, 30, 60 min 

Internal memory 

Circular management or stop logging if memory is full. 

The number of storable samples depends on the number of detected 

quantities (see table 9.10.1). 

Alarm 

Acoustic through internal buzzer 

Display 

Optional custom LCD  

Buttons 

Connection button inside the instrument 

LED indicators 

RF communication status. Models without LCD have alarm and battery 

level LEDs. 

Power supply 

Non rechargeable lithium thyonil chloride (Li-SOCl

2

) internal battery, 3.6 

V, size A (size C for HD35EDWK/4TC, HD35EDWS/xTC and HD35EDWH), 

2-pole Molex 5264 connector. Optional 24 Vac/dc power supply. 

7…30 Vdc external power supply (without internal battery) for HD35EDWPTC-

ALM, HD35EDW-MB and the models with CO

2

 probe. 

Battery life 
(without repeaters, 

direct communication 

with  HD35AP...) 

4 years typical for HD35EDWK/4 and HD35EDWH models (with 10 s meas-

urement interval and 30 s logging interval); 

2 years typical for the other models, with 5 s measurement interval (10 s 

for HD35EDW7P/...TC, HD35EDW14bNTC, HD35EDW14b7PTC, 

HD35EDWWBGT) and 30 s logging interval. 

Operating temperature 

and humidity  

-20…+70 °C / 0…100

 

%RH 

Dimensions 

See dimensional drawing  

Connectors for external 

probes with cable 

Depending on the model: M12 connectors, thermocouple connectors or

 

terminal

 

header

 

inputs

 

3.5

 

mm pitch. 

Weight 

250 g approx. (battery included) 

Housing 

Polycarbonate 

Protection Degree 

IP 67 (IP65 for the models with CO

2

 sensor) 

Installation 

Wall mount or fixing to a 40mm diameter mast through the clamping 

HD2003.77/40 

(optional, for housing H=120 mm x L=80 mm). 

Optional protection shield from solar radiations. 

(*)

 Some models that measure many quantities may have a minimum interval higher than 1 second (see 

table 9.10.1). 

Содержание HD35AP series

Страница 1: ...Operating manual Wireless data logging HD35 series www deltaohm com Companies Brands of GHM English Keep for future reference...

Страница 2: ...TECHNICAL FEATURES OF BASE UNIT HD35APR 29 6 12 TECHNICAL FEATURES OF BASE UNITS HD35AP GMT 30 7 HD35RE REPEATERS 31 7 1 AVAILABLE VERSIONS 31 7 2 DESCRIPTION OF HD35RE IN HOUSING FOR INDOOR USE 31 7...

Страница 3: ...DER INPUTS 76 9 6 CONNECTING THE MODEL HD35EDW MB WITH RS485 MODBUS RTU INPUT 78 9 7 ADAPTER FOR CO2 CALIBRATION WITH THE CYLINDER 79 9 8 RAIN GAUGE CONNECTION 79 9 9 WBGT INDEX MEASUREMENT 80 9 10 TE...

Страница 4: ...tion to the ones mentioned above For each detected quantity the user can set two alarm thresholds high threshold and low thresh old The alarm hysteresis and delay can be configured for each quantity T...

Страница 5: ...The data transferred to the PC are entered into a database Data loggers comply with EN 12830 standard The optional HD35AP CFR21 advanced ver sion of the software is designed in compliance with FDA 21...

Страница 6: ...s HD35ED allowing the communication distance between data loggers and base unit to be increased Several repeaters in cascade can be used Fig 2 2 repeater placed between data loggers and base unit HD35...

Страница 7: ...ure 1 Install HD35AP S and MySQL softwares 2 Connect the battery to the base unit 3 Connect the base unit to the PC 4 Connect the software to the base unit 8 Add devices to the network 9 Check RF comm...

Страница 8: ...housing by fixing the 4 front screws Factory configured system first connect the battery of the base unit and check the clock setting Only after this operation connect the battery of the data loggers...

Страница 9: ...data loggers synchronize the clock with that of the base unit If the base unit is not yet connected to the power supply the synchronization fails and data loggers with LCD option display the clock set...

Страница 10: ...NETWORK If the network structure is not configured it is necessary to add data loggers and any repeat ers and alarm modules to the base unit The adding procedure described in detail in the chap ter A...

Страница 11: ...ER Packet Er ror Rate is close to zero Fig 3 7 verification of RF communication with HD35AP S software In data loggers with display by checking that the strength of the received signal RSSI exceeds 85...

Страница 12: ...02 928 U S A and Canada 915 928 Australia or 921 5 928 New Zealand MHz frequency band To change the band in the devices with LCD select the RF Frequency item of the RF Parameters menu models with grap...

Страница 13: ...ail in the chapter Removing devices from the network of the software online help and briefly described here under 1 Select the command Network of the HD35AP S software 2 In the section Delete Devices...

Страница 14: ...possibility of applying a padlock inserted in a fastening pin to be fixed to the wall Removable installation 1 Fix the plastic support to a wall 2 Insert the device into the support pushing it downwar...

Страница 15: ...feature multiple PCs can be connected simultaneously via TCP IP to the same base unit HD35APG With USB output and integrated GSM GPRS module for e mail or SMS alarm transmission and stored data trans...

Страница 16: ...cy the blink period increases of 1 second for each 10 decrease of the battery charge 4 Connector for external 6 Vdc power supply SWD06 5 Mini USB connector for PC connection The PC USB port powers the...

Страница 17: ...ly 4 RS485 Modbus RTU protocol connector 5 ETHERNET RJ45 connector 6 Mini USB connector for PC connection The PC USB port powers the instrument in the ab sence of an external power supply 7 Bicolor RF...

Страница 18: ...F antenna for transmission in ISM band 2 Input for external power supply 3 Mini USB connector for PC connection 4 Housing closing hooks 5 GSM 3G antenna Place the GSM 3G antenna at least 30 cm away fr...

Страница 19: ...ed to a PC through the mini USB connector and CP31 cable di rectly with type A USB connector for HD35APD In this connection mode the base unit is powered through the PC USB port Fig 6 4 1 USB connecti...

Страница 20: ...s a dynamic IP address to the network server router Fig 6 5 2 base unit in Wi Fi mode operating as a client It is possible to access the base unit from any PC of the local network where the basic HD35...

Страница 21: ...he ends of the network The cable shield must be connect ed to both ends of the line The maximum number of devices that can be connected to the line Bus RS485 depends on the load characteristics of the...

Страница 22: ...the RS485 line insert the termination by placing the short jumper next to the RS485 connector between the RT and CLOSE indications Otherwise remove the termination by placing the short jumper between...

Страница 23: ...t the SIM tray back in place making sure that the SIM card contacts face the instrument board 6 Reconnect the battery 7 Close again the housing by fixing the 4 front screws Through the HD35AP S softwa...

Страница 24: ...ur 3 2 hours 4 4 hours 5 8 hours 6 12 hours 7 24 hours 8 2 days 9 4 days 10 1 week FTP FORMAT format index Set the format of the data sent via FTP where format index 1 log format for database 2 csv fo...

Страница 25: ...tes immediate data upload on the HTTP server starting from the last measurement transmitted HTTP DL FROM YYYY MM DD HH MM SS Uploads data on the HTTP server starting from the specified date where YYYY...

Страница 26: ...on request The server IP address HD35AP G must be public and static Connection HD35AP G Client PC Server 1 Open a port port forwarding in the Modem Router through which your PC connects to In ternet f...

Страница 27: ...o HD35AP G the SMS command TCP SERVER ON 3 HD35AP G replies with a first SMS to confirm that the command has been accepted Wait for a second SMS with the confirmation that the TCP server functionality...

Страница 28: ...e mail to an FTP address or to an HTTP server Cloud Allows GPRS 3G TCP IP protocol Internal memory The number of storable samples depends on the type of data loggers con nected The capacity is of 226...

Страница 29: ...r 915 9 929 7 MHz depending on the model Antenna Whip external Transmitting range In open field 300 m E J 180 m U towards data loggers with internal antenna 500 m E J U towards repeaters and data logg...

Страница 30: ...onnector cable CP31 Ethernet connection No Wi Fi connection No GSM GPRS connection Yes for sending alarm e mail or SMS and data via e mail to an FTP address or to an HTTP server Cloud Allows the GPRS...

Страница 31: ...s fully charged as the battery is running low the LED blinks with a lower and lower frequency the blink period increases of 1 second for each 10 de crease of the battery charge 4 Connector for externa...

Страница 32: ...condition 7 4 CONNECTION TO THE WIRELESS NETWORK The device can be connected and disconnected to from the wireless network by pressing for 5 seconds o the connection button on the front panel for HD35...

Страница 33: ...s can be interposed as shown in following examples Fig 7 5 1 arrangement of different types of repeaters Unlike HD35RE repeaters which have external power supply the HD35REW repeaters are powered only...

Страница 34: ...ble Only for configuration and firmware update not for data down load LED indicators Presence of external power supply battery charge level RF communica tion status Keyboard Connection PING for testin...

Страница 35: ...ic conditions LED indicators Battery charge level RF communication status Push buttons Push button for connection inside the instrument Power supply Internal 3 6 V lithium thionyl chloride Li SOCl2 no...

Страница 36: ...l see the following page 6 Connection PING for testing RF button 7 Internal RF antenna 8 MENU ENTER key allows access to the configuration menu confirm the selected option or the set value in the menu...

Страница 37: ...ectable probes HD35ED7P 2TC TP35 Pt100 Pt1000 HD35EDN 2TC TP35N NTC HD35ED1N 2TC IN 1 HP3517TC TP35N NTC IN 2 TP35N NTC Model Connectable probes HD35ED7P 3TC TP35 Pt100 Pt1000 HD35EDN 3TC TP35N NTC M1...

Страница 38: ...1NUBTCV LP35UVB HD35ED1NUCTCV LP35UVC HD35ED14bNIUTCV LP35P A Fixed vertical RH T probe Differential pressure inputs metal inputs with tube clamp ring in HD35ED1N4r5TV Ingressi pressione differenziale...

Страница 39: ...eries with NTC 10K 25 C or Pt100 temperature sensor accord ing to the model Replacement of the HP3517 probe requires recalibration of the in strument in line with the new probe In models with 2 or 3 M...

Страница 40: ...D HD35ED 1 TVI D HD35ED 1N TC 1 A HD35ED 17P TC 1 A HD35ED 1N TV D HD35ED 1N TVI Sensor integrated in RH module D HD35ED 1N 2 TC 2 B HD35ED 1N 2 TCV 1 T RH I HD35ED 14bN TC 1 Patm A HD35ED 14bN TV D...

Страница 41: ...Custom Graphic HD35ED 1NB Sensor integrated in RH module F HD35ED 1NAB F HD35ED 14bNAB F HD35ED H Transmitters with 0 20 mA 4 20 mA 0 50 mV or 0 1 V output Pt100 Pt1000 Sensors K J T N E thermocouples...

Страница 42: ...t connected the icon has the aspect of a hang up phone in data loggers with graphic LCD and it is off in data loggers wit custom LCD PING function In the devices connected to a wireless network it is...

Страница 43: ...point Atmospheric pressure PRES alternated to m u Atmospheric pressure Atm Pres Differential pressure PRES alternated to m u Carbon monoxide Carbon monoxide CO Carbon dioxide Carbon dioxide CO2 Illum...

Страница 44: ...Entering the user password makes some settings not changeable The instrument exits automatically the menu if no key is pressed for 3 minutes After exiting the menu the password remains active for a f...

Страница 45: ...that particular model 1 Information It lists the general information of the instrument model serial number RF address user code firmware version calibration date etc 2 Display configuration 1 Measure...

Страница 46: ...he main menu 5 Alarm thresholds or Alarm configuration 1 Quantity 1 down threshold lower alarm threshold of quantity 1 The type of quantity depends on the data logger model 2 Quantity 1 up threshold h...

Страница 47: ...nly the temperature measurement unit is available Meas urement units for the other quantities are configurable in the menus for the setting of channels Ch 1 Ch 2 e Ch 3 Note the unit of measurement is...

Страница 48: ...3 Auto cal 1st period time interval after which the first auto calibration will be performed after activation 4 Background CO2 value CO2 reference value for auto calibration 5 Auto cal max change maxi...

Страница 49: ...tructure of the main menu with the relevant submenus is shown below Accord ing to the data logger model some items could be not available if not significant for that par ticular model 1 DEV_INFO infor...

Страница 50: ...the data logger model 5 Quantity n_UP_THLD higher alarm threshold of quantity n The type of quantity depends on the data logger model 6 THLD_ALRM enables or disables the buzzer when measurement thresh...

Страница 51: ...on cyclical management logging stops when the memory is full of the data logger memory Select YES for the cyclical management 3 LOG RF_TIME choice of logging and RF transmission interval the two inter...

Страница 52: ...strument to the pressure measuring point default 6 mm The item appears only in HD35ED 4r5TV models 5 ABS_PRES_mbar absolute pressure in mbar of the air in the connecting tube The item appears only in...

Страница 53: ...n the range 0 to 20 mA 8 5 1 INPUTS CONFIGURATION Configuration of inputs can be implemented through the HD35AP S software see software in structions or if the data logger is equipped with a display t...

Страница 54: ...on OK and confirm with ENTER to continue 4 A message will be displayed reminding that the two coordinates of the linear relation be tween input and physical quantity will be now required x1 input valu...

Страница 55: ...he User option The calibration procedure deletes the data of the previous user calibration In case of failed procedure you can always return the instrument to factory calibration by selecting o in mod...

Страница 56: ...ion grid by fixing it with the two screws 8 6 2 REPLACING THE CO SENSOR The CO sensor has an expected average life of over 5 years of use in normal conditions Where it becomes necessary to replace the...

Страница 57: ...alibrated to any reference value within the measurement range Calibration procedure 1 Place the instrument in an environment with known CO2 concentration for ex in clean air 2 Wait for at least 15 min...

Страница 58: ...known value referred to as environment back ground value For example we can have that an instrument installed inside a public place performs a weekly auto calibration when people are not present and...

Страница 59: ...the M12 1 threaded ring Avoid any contact of the sensitive element with your hands or other objects or liquids If some liquid has formed inside the measuring chamber dry it with a clean absor bent pap...

Страница 60: ...he instrument pressure input open 2 Access menu with the administrator password and select the item CAL_MENU DIFF_PRES_0 Pa_CAL Press ENTER to confirm 3 The zero value blinks on the display DIFF 0 LOG...

Страница 61: ...uantities see table 8 7 1 Alarm Acoustic by means of the internal buzzer Power supply Non rechargeable lithium thyonil chloride Li SOCl2 internal battery 3 6 V size A 2 pole Molex 5264 connector Batte...

Страница 62: ...CV 44 000 1 s T RH TD TW AH MR PVP UVC HD35ED 14bNIU TCV 32 000 2 s T RH TD TW AH MR PVP PATM I UVA PUV HD35ED 14bNIU TV 32 000 2 s T RH TD TW AH MR PVP PATM I UVA PUV HD35ED 1NB 44 000 10 s T RH TD T...

Страница 63: ...max for probes measuring only temperature the measuring range can be limited by the operating temperature of the used probe 40 150 C for T RH combined probes HD3517ETC Resolution instrument 0 1 C Acc...

Страница 64: ...eter Carbon Monoxide CO Sensor Electrochemical cell Measuring range 0 500 ppm Resolution instrument 1 ppm Accuracy 3 ppm 3 of measurement Operating temperature 5 50 C Response time T90 50 s Stability...

Страница 65: ...e See graph 4 Calibration uncertainty 5 f2 response as cosine law 6 f3 linearity 1 f4 instrument reading error 1 digit f5 fatigue 0 5 One year drift 2 Operating temperature 0 50 C Illuminance Sensor P...

Страница 66: ...UVA irradiance sensor Graph 3 Relative spectral response of UVB irradiance sensor Graph 4 Relative spectral response of UVC irradiance sensor WAVE LENGTH nm RELATIVE SPECTRAL RESPONSE WAVE LENGTH nm R...

Страница 67: ...e J 100 750 C type E 200 750 C type T 200 400 C type N 200 1300 C Resolution 0 1 C Accuracy excluding probe error type K 0 1 C 600 C type E 0 1 C 300 C 0 2 C 600 C 0 2 C 300 C type N 0 1 C 600 C type...

Страница 68: ...ttery As the bat tery runs low the LED blinks with a lower and lower frequency the blink period increases of 1 second for each 10 decrease of the battery charge 3 ALARM LED red color it blinks when th...

Страница 69: ...P 2TC TP35 Pt100 Pt1000 HD35EDWN 2TC TP35N NTC HD35EDW1N 2TC IN 1 HP3517TC TP35N NTC IN 2 TP35N NTC HD35EDW1NRTC IN 1 HP3517TC TP35N NTC IN 2 LP PYRA 02 LP PYRA 03 LP SILICON PYRA 04 HD35EDW7PRTC IN 1...

Страница 70: ...5EDWS 3TC HP3510 1 HP3510 2 HD35EDWWBGT IN 1 TP3507TC2 IN 2 TP3575TC2 TP3576TC2 globe thermometer IN 3 TP3501TC2 wet bulb TP3204 wet bulb Model Connectable probes HD35EDW1NI2TCV LP35PHOT03BL M12 Conne...

Страница 71: ...nectable probes HD35EDWPTC ALM HD2013 HD2015 Model Connectable probes HD35EDWDPTC IN 1 HP712 IN 2 HD2013 HD2015 M12 Connector Vertical fixed RH T probe CO2 probe Cable gland for power supply USB Alarm...

Страница 72: ...ange 0 10 000 ppm TC Probe with cable TV or TVI Fixed vertical probe without cable TCV Probe with cable fixed vertical probe without cable Models measuring temperature and humidity with a combined pro...

Страница 73: ...C 2 B HD35EDW 1NP TC 2 B HD35EDW 1NL TC 2 B HD35EDW S TC Soil temperature and volumetric water content 1 A HD35EDW S 2 TC 2 B HD35EDW S 3 TC 3 C HD35EDW DP TC Level 1 cable gland M HD35EDW 1NI2 TCV Lu...

Страница 74: ...ature and relative humidity probe requires HD9007A 1 or HD9007A 2 protection against solar radiations 9 3 CONNECTION TO THE WIRELESS NETWORK The device can be connected and disconnected to from the wi...

Страница 75: ...e data logger is not connected PING function In the devices connected to a wireless network it is possible to check if the base unit can be reached by briefly pressing the connection button if the gre...

Страница 76: ...erminals of the channel to 0 10V In all other cases set the switch to 0 1V If a channel is configured as current input insert the 50 shunt resistance closing the jumper placed above the terminals of t...

Страница 77: ...he version for 7 28 Vdc external power supply connect the external power supply to the terminals shown in Fig 9 5 4 Fig 9 5 3 connection of the battery in the model HD35ED L WH Fig 9 5 4 connection of...

Страница 78: ...1 connection of HD35EDW MB model The jumper placed above the terminals allows inserting the termination resistance for long dis tance RS485 connection Place the jumper between the RT and CLOSE indicat...

Страница 79: ...inder unscrew the probe filter screw the HD31 B3A adapter and connect the cylinder adjust the bottle flow meter to get a constant flow between 0 3 and 0 5 l min Fig 9 7 1 CO2 calibration with cylinder...

Страница 80: ...tted to in a warm environment For WBGT index measurement refer to ISO 7243 standard To calculate WBGT index according to ISO 7243 standard it s necessary that to the instrument are connected TP3501TC2...

Страница 81: ...obe The cotton wick must protrude from the probe for about 20 mm 3 Replace the cap 4 Fill the bottle with 500 cc of distilled water 5 Turn the probe over and firmly screw the bottle to the probe reser...

Страница 82: ...meter TP3204 wet bulb probe HD35EDWWBGT data logger HD32 2 7 1 support HD35 24W flange Screw for fixing the HD32 2 7 1 support to the tripod TP3204 wet bulb probe TP3575TC2 or TP3576TC2 globe thermome...

Страница 83: ...tive humidity probe Pt100 temperature sensor The WBGT index measurement with combined temperature and relative humidity probe is not according to ISO 7243 The outdoor installation of the HP3517ETC2 pr...

Страница 84: ...and battery level LEDs Power supply Non rechargeable lithium thyonil chloride Li SOCl2 internal battery 3 6 V size A size C for HD35EDWK 4TC HD35EDWS xTC and HD35EDWH 2 pole Molex 5264 connector Optio...

Страница 85: ...0 1 s T VWC HD35EDW S 2 TC 36 000 1 s T VWC HD35EDW S 3 TC 26 000 1 s T VWC HD35EDW DP TC 28 000 1 s FL PREL P DP IP HD35EDW 1NI2 TCV 30 000 1 s T RH TD AH I HD35EDW 1NB TV 30 000 1 s T RH TD AH CO2 H...

Страница 86: ...tegrated in RH module For versions TVI Sensor Sensor integrated in humidity module Measurement range 40 105 C Resolution of instrument 0 1 C Accuracy 0 2 C in the range 0 60 C 0 2 0 05 T C in the rang...

Страница 87: ...0 1 C year Relative Humidity For versions TC and TV Sensor Capacitive the sensor can be calibrated by user Measurement range 0 100 RH Resolution of instrument 0 1 Accuracy 1 8 RH 0 85 RH 2 5 RH 85 100...

Страница 88: ...or of pressure relative to the atmosphere Pressure measuring range 0 1 bar Level measuring range Depends on the fluid density configurable via software For water 0 10 m approx Resolution of the instru...

Страница 89: ...photopic re sponse V 6 f2 response as cosine law 3 f3 linearity 1 f4 instrument reading error 0 5 f5 fatigue 0 5 Class B One year drift 1 Operating temperature 0 50 C Reference standard CIE n 69 UNI 1...

Страница 90: ...J 100 750 C type E 200 750 C type T 200 400 C type N 200 1300 C Resolution 0 1 C Accuracy excluding probe error type K 0 1 C 600 C type E 0 1 C 300 C 0 2 C 600 C 0 2 C 300 C type N 0 1 C 600 C type J...

Страница 91: ...auge Cup anemometer Wind direction vane Only one probe of each type can be connected It is also possible to connect only some of the probes 2 Connection button 3 External RF Antenna optional The anten...

Страница 92: ...models provided with input for the solar radiation sensor option R the pyranometers LP PYRA 02 LP PYRA 03 LP SILICON PYRA 04 etc with mV output signal can be connected In addition to the solar radiati...

Страница 93: ...ing for 5 seconds the connection button the buzzer emits a beep and the RF LED blinks green for 1 second to indicate the start of the connection proce dure If the device belongs to a wireless network...

Страница 94: ...5 DISPLAY Through HD35AP S software you can select the measurement to be displayed on the main line of the display or set the automatic alternation of measured quantities Information on the con nectio...

Страница 95: ...samples from 28 000 to 58 000 depending on the number of detected quantities Alarm Acoustic through internal buzzer Display Optional custom LCD Buttons Watertight push button for connection PING for t...

Страница 96: ...1 hPa 300 1100 hPa T 0 50 C Stability 1 hPa anno Temperature drift 3 hPa tra 20 60 C Solar radiation Sensor Thermopile Measuring range 0 2000 W m2 Resolution of the instrument 1 W m2 Sensitivity Confi...

Страница 97: ...installed on a flat terrain site Offset 0 35 m s Gain 0 765 m s 1 Hz Distance constant 63 recovery 2 55 m 5 m s 2 56 m 10 m s ASTM D 5096 02 Calculated quantities Felt air temperature as a function of...

Страница 98: ...are protected by a cover 4 Green RF LED it blinks if RF transmission was successful 5 Red RF LED it blinks to indicate that RF transmission has failed 6 Connection PING for testing RF button 7 Interna...

Страница 99: ...e otherwise it will be the red RF LED to blink 11 4 TECHNICAL CHARACTERISTICS OF THE ALARM DEVICE Transmission Frequency 868 MHz 902 928 MHz or 915 9 929 7 MHz according to the model Antenna Internal...

Страница 100: ...e 32767 In case of doubt on the registers actually available in a particular model use the function Download the list of MODBUS registers of the device included in the Settings sections of HD35AP S so...

Страница 101: ...se of measurement alarm 0 no 1 yes 5 b If 1 there are unsaved device parameters in the flash memory Set 0 to force storage 6 b If 1 there is a pending RF rescheduling RF transmission se quence of devi...

Страница 102: ...t measurement unit x10 13 B Wet bulb temperature alarm 0 OFF 1 lower threshold alarm 2 higher threshold alarm 16 SW TEMPERATURE with NTC10K sensor of channel 2 in the set measurement unit x10 17 B Ala...

Страница 103: ...3 0 OFF 1 lower threshold alarm 2 higher threshold alarm 58 SW WIND CHILL in the set measurement unit x10 59 B Wind chill alarm 0 OFF 1 lower threshold alarm 2 higher threshold alarm 60 SW DIFFERENTIA...

Страница 104: ...d alarm 2 higher threshold alarm 92 SW WIND SPEED HD52 3D anemometer in m s x100 93 B Wind speed HD52 3D anemometer alarm 0 OFF 1 lower threshold alarm 2 higher threshold alarm 94 SW WIND DIRECTION HD...

Страница 105: ...reshold alarm 1002 200x N 1 SW TEMPERATURE with 3 wire Pt100 sensor of channel N in the set measurement unit x10 1003 200x N 1 B Alarm for temperature with 3 wire Pt100 sensor of channel N 0 OFF 1 low...

Страница 106: ...1034 200x N 1 SW Position of potentiometer in of channel N Only if channel N is configured as potentiometric input HD35EDH 1035 200x N 1 B Alarm for channel N if the channel is configured as potentio...

Страница 107: ...INANCE resolution 2 100 1 10 0 1 5024 W ATMOSPHERIC PRESSURE measurement unit see TAB 12 1 5025 SW ATMOSPHERIC PRESSURE resolution 2 100 1 10 0 1 1 0 1 2 0 01 5046 W Unit of measurement for TEMPERATUR...

Страница 108: ...input HD35EDH 2 100 1 10 0 1 1 0 1 2 0 01 6038 200x N 1 W Measurement unit of the quantity associated to channel N if the channel is configured as 0 50 mV HD35EDH See TAB 12 1 6039 200x N 1 SW Resolu...

Страница 109: ...ermined 10019 W Alarm relay 2 status 0 deactivated 1 intermittent 2 active 3 undetermined TAB 12 5 Holding Registers Read Write parameters Address Type Holding Register description AP ED RE AL Measure...

Страница 110: ...threshold for VWC PROBE OUTPUT in mV x10 37 SW Higher alarm threshold for VWC probe output in mV x10 46 SW Lower alarm threshold for TEMPERATURE with sensor integrated in RH module in the set measure...

Страница 111: ...r alarm threshold for WET BULB TEMPERATURE measured by the natural ventilation wet bulb probe in the set measurement unit x10 73 SW Higher alarm threshold for wet bulb temperature measured by the natu...

Страница 112: ...er alarm threshold in l s 106 SW FLOW lower alarm threshold in l min 107 SW Flow higher alarm threshold in l min 108 SW FLOW lower alarm threshold in m3 min 109 SW Flow higher alarm threshold in m3 mi...

Страница 113: ...er alarm threshold for TEMPERATURE with TC_K sensor of channel N in the set measurement unit x10 1013 200x N 1 SW Higher alarm threshold for temperature with TC_K sensor of channel N in the set measur...

Страница 114: ...when the channel is configured as potenti ometric input HD35EDH 1043 200x N 1 SW Higher alarm threshold expressed as value of the quantity asso ciated to channel N when the channel is configured as p...

Страница 115: ...The setting is extended to all EDs except for the mapped quantities in HD35EDH 10049 W Differential pressure measurement unit for ranges r1 r2 and r3 see TAB 12 1 The setting is extended to all EDs ex...

Страница 116: ...with ASCII codification of measurement 6 Available for models with more measurements of the same type 20072 to 20083 B User code with ASCII codification of measurement 7 Available for models with more...

Страница 117: ...following figures show the dimensions of the instruments in mm HD35AP HD35APW HD35APG HD35RE HD35APS HD35ED versions with fixed RH T probe HD35ED versions with grid HD35ED versions with M12 connector...

Страница 118: ...HD35 118 V1 12 HD35ED versions with Support for removable installation differential pressure inputs Flanges for fixed installation Lock pin HD35APR...

Страница 119: ...HD35 119 V1 12 HD35REW HD35EDW versions with M12 connectors HD35EDW versions with fixed RH T probe HD35EDW versions with terminal header HD35EDW versions with CO2 probe...

Страница 120: ...HD35 120 V1 12 HD35EDLM TC...

Страница 121: ...tandard 40 150 C with option E Accuracy 1 8 RH 0 85 RH 2 5 RH 85 100 RH T 15 35 C 2 1 5 measure T remaining range Cable length 2 5 or 10 m standard Connection 4 pole HD3517TC or 8 pole HP3517ETC M12 f...

Страница 122: ...ess steel tube TP472I Sensor 4 wire Pt1000 Temperature range 196 500 C Cable length 2 m Connection Free wires option W or 4 pole M12 female connector option C Material AISI 316 stainless steel tube TP...

Страница 123: ...ree wires or 4 pole M12 female connector option C Material AISI 316 stainless steel tube TP35 5AF1 2 Sensor 4 wire Pt100 Temperature range 110 180 C Accuracy Class A Cable length 2 m standard other le...

Страница 124: ...nless steel tube TP35N2 Sensor NTC 10 k 25 C Temperature range 0 75 C Accuracy 0 3 C Cable length 3 5 o 10 m standard other lengths on request Connection Free wires or 4 pole M12 female connector opti...

Страница 125: ...th K thermocouple sensor TP35K6 5 Sensor K thermocouple with isolated junction Max temperature 50 750 C Accuracy Class 1 according to IEC 60584 2 Cable length 5 m standard other lengths on request Con...

Страница 126: ...ctions of Pt100 Pt1000 and NTC 10K temperature probes with M12 connector Reference External view of the instrument M12 connector Probe M12 connector NTC 10K Pt100 Pt1000 4 wires Pt100 Pt1000 2 wires P...

Страница 127: ...fic measures described in this manual are followed The instruments proper operation and operating safety can be ensured only in the climatic conditions specified in this manual Do not use the instrume...

Страница 128: ...GMT and HD35AP3GMT internal rechargeable battery wall mount support HD35 03 not for HD35APD HD35APR HD35APGMT and HD35AP3GMT operating manual The radio frequency 868 902 928 or 915 9 929 7 MHz has to...

Страница 129: ...wo relay outputs for signaling alarm events Power supply through non rechargeable lithium thionyl chloride Li SOCl2 3 6V internal battery bat tery life 1 year typical Supplied with battery HD35 BAT2 w...

Страница 130: ...nthetically Active Radiation P rainfall quantity R solar radiation pyranometer U UV irradiance U UVA UB UVB UC UVC S soil volumetric water content and temperature L leaf wetness D level W wind speed c...

Страница 131: ...oplastic rubber temperature probe 3 wire Pt1000 1 3 DIN sensor Di mensions 5 x 20 mm Cable length 3 or 5 m standard Cable ending with free wires or 4 pole M12 female connector Operating temperature 0...

Страница 132: ...or 10 m standard 4 pole M12 fe male connector Operating temperature 40 105 C TP35N1 Stainless steel temperature probe NTC 10K 25 C sensor Dimensions 5 x 40 mm Cable length 3 5 or 10 m standard Cable...

Страница 133: ...BGT measurement system to a wall or a mast HD2013 2 14 3 sector clamping for mast 40 mm with 6 inputs 16 mm HD3218K Clamp shaft for fixing a probe to the HD2013 2 14 flange PHOTOMETRIC RADIOMETRIC PRO...

Страница 134: ...or mast 40 50 mm LP S3 Attachment bracket for LP PYRA 03 suitable for mast 40 50 mm RAIN GAUGES HD2013 Rain gauge with tipping bucket area 400 cm2 for temperature range 4 C 60 C Standard resolution 0...

Страница 135: ...or WIND SPEED AND DIRECTION SENSORS HD54 3 Passive cup anemometer Measuring range 1 75 m s Operating condi tions 45 60 C 0 100 RH Rod mounting Height 81 mm assembled HD54 D Wind direction vane probe M...

Страница 136: ...ic water content With inte grated NTC 10 k temperature sensor M12 connector 10 m cable HP3510 2 5 3 electrode probe for measuring the soil volumetric water content in restricted volumes With integrate...

Страница 137: ...ole M12 connector on one side free wires on the other Length 20 m CPM12 AA4 2D Cable with 4 pole M12 connector on both sides Length 2 m CPM12 AA4 5D Cable with 4 pole M12 connector on both sides Lengt...

Страница 138: ...0 mm tube 6 inputs 16 mm HD2005 20 Anodized aluminum tripod kit with adjustable legs for installing environ mental sensors Height 2 m It can be fixed on a flat base with screws or to the ground with p...

Страница 139: ...i certifications Models HD35APW E and HD35APW U contain IEEE 802 11b g n module FCC ID XM5 SMG2SMT IC ID 8516A SMG2SMT Model HD35APW J contains IEEE 802 11b g n module FCC ID XM5 SMG2N2 IC ID 8516A SM...

Страница 140: ...rent from that to which the receiver is connected Consult the dealer or an experienced radio TV technician for help Notice To satisfy FCC IC RF exposure requirements for mobile and base station transm...

Страница 141: ...NOTES...

Страница 142: ...NOTES...

Страница 143: ...NOTES...

Страница 144: ...d third parties is excluded from the guarantee The instrument shall be returned FREE OF SHIPMENT CHARGES to your dealer The jurisdiction of Padua applies in any dispute The electrical and electronic e...

Страница 145: ...e result of the constant development of the product This may produce some differences between the information written in this manual and the instrument you have purchased We cannot completely exclude...

Страница 146: ...GHM GROUP Delta OHM Delta Ohm S r l a socio unico Via Marconi 5 35030 Caselle di Selvazzano Padova ITALY Phone 39 049 8977150 Fax 39 049 635596 www deltaohm com info deltaohm com V1 12 10 01 2018...

Отзывы: