HD35
- 109 -
V1.12
Address Type
Input Register description
AP ED RE AL
10002
W
Day of last measurement.
10003
W
Hour of last measurement.
10004
W
Minutes of last measurement.
10005
W
Seconds of last measurement.
10006
W
Packet Error Rate in % of the device (x10).
10007
W
Number of RF hops of the last transmitted packet.
10008
SW RF signal level in dBm (relating to last RF hop).
10009
W
Battery level:
0=empty, 1=half full , 2=full, 3=external power supply
10010
W
Time, in seconds, elapsed since the last transmitted packet.
10011
W
RF signal level expressed as 0 to 7 scale.
10012
W
Modbus address of the AP to which the device is connected.
10013
W
Password level for the current connection:
0=no password, 1=user level, 2= administrator level
10014
W
Battery remaining capacity in %.
10015
W
Estimation of battery remaining capacity in hours (x10).
10016
W
Type of power supply: 0=battery, 1=USB, 2=ext. power supply
10017
W
Estimation of battery remaining capacity in weeks
10018
W
Alarm relay #1 status:
0=deactivated, 1=intermittent, 2=active, 3=undetermined
10019
W
Alarm relay #2 status:
0=deactivated, 1=intermittent, 2=active, 3=undetermined
TAB. 12.5: Holding Registers – Read/Write parameters
Address Type
Holding Register description
AP ED RE AL
Measurement alarm thresholds
0
SW Lower alarm threshold for
TEMPERATURE
with NTC10K sensor of
channel 1 in the set measurement unit (x10).
1
SW Higher alarm threshold for temperature with NTC10K sensor of
channel 1 in the set measurement unit (x10).
2
SW RH lower alarm threshold in % (x10). Only for models …TC
and …TV.
3
SW RH higher alarm threshold in % (x10). Only for models …TC
and …TV.
4
SW D
EW POINT
lower alarm threshold in the set measurement unit
(x10).
5
SW Dew point higher alarm threshold in the set measurement unit
(x10).
6
SW P
ARTIAL VAPOR PRESSURE
lower alarm threshold in hPa (x100).
7
SW Partial vapor pressure higher alarm threshold in hPa (x100).
8
SW M
IXING RATIO
lower alarm threshold in g/Kg (x10).
9
SW Mixing ratio higher alarm threshold in g/Kg (x10).
10
SW A
BSOLUTE HUMIDITY
lower alarm threshold in g/m
3
(x10).
11
SW Absolute humidity higher alarm threshold in g/m
3
(x10).
12
SW W
ET BULB TEMPERATURE
lower alarm threshold in the set measure-
ment unit (x10).
13
SW Wet bulb temperature higher alarm threshold in the set measure-
ment unit (x10).