HD35
- 81 -
V1.12
Warning: don't turn the probe from the vertical direction because the distilled water can exit.
The cotton wick calcifies (becomes hard) with time: replace it periodically.
Preparation of the TP3204 wet bulb temperature probe:
1.
Remove the sensor cap (the cap is not screwed).
2.
Insert the cotton wick, previously dipped with distilled water, into the temperature probe.
The cotton wick must protrude from the probe for about 20 mm.
3.
Replace the cap.
4.
Fill the bottle with 500 cc of distilled water.
5.
Turn the probe over and firmly screw the bottle to the probe reservoir.
6.
Turn the probe quickly (to avoid water spillage).
7.
Secure the probe to the HD32.2.7.1 support by using the two screws at the bottom of the
probe.
For measurements in presence of solar irradiation, use the protection shield from solar radiations.
The cotton wick calcifies (becomes hard) with time: replace it periodically.
Screws
Reservoir
Cap
Cotton wick
Shield
Bottle