HD35
- 76 -
V1.12
9.5
C
ONNECTING THE MODEL
HD35EDWH
WITH TERMINAL HEADER INPUTS
HD35ED[L]WH model has four terminal header inputs. Each input can be configured as a
Pt100/Pt1000, thermocouple, 0/4…20 mA (shunt resistance inside), 0…50 mV, 0…1 V, 0…10 V or
potentiometric input. Only input 4 can be also configured as pulse counter (count of voltage-free
contact switchings).
Fig. 9.5.1: sensor connection in HD35EDWH model
If a channel is configured as 0-10 V input, set the switch on the terminals of the channel to
0-10V. In all other cases, set the switch to 0-1V.
If a channel is configured as current input, insert the 50
Ω
shunt resistance closing the jumper
placed above the terminals of the relevant channel. In all the other configurations, leave the
jumper open.
Fig. 9.5.2: shunt resistance for current input
The current input accepts any value in the range 0 to 20 mA.
Configuration of inputs is done with the HD35AP-S software (see software instructions).
Input 1
Input 4
Input 2
Input 3
Pt100
Pt1000
2 wires
0-20 mA
4-20 mA
ON/OFF
contact
Pt100
Pt1000
3 wires
Pt100
Pt1000
4 wires
Thermocouple
K,J,T,N,E
0-1 V
Potenziometer
0-50 mV
0-10 V
Jumper
Shunt
resistance