HD35
- 61 -
V1.12
8.7
T
ECHNICAL
C
HARACTERISTICS OF DATA LOGGERS IN INDOOR USE HOUSING
Transmission frequency
868 MHz, 902-928 MHz or 915.9-929.7 MHz according to the model
Antenna
Internal
Transmission range
In open field:
300 m (E, J)/ 180 m (U) towards base unit (except HD35APD...) and re-
peaters.
180 m (E, U) towards base unit HD35APD.
(could be reduced if obstacles or adverse weather conditions are
present)
Display
Optional. Custom or graphic LCD according to the model (see table 8.1.2).
Keyboard
Connection / PING (for testing RF) button.
Models with LCD have buttons for the configuration and scrolling of
measured values.
LED Indicators
RF communication status. Models without LCD have alarm and battery
level LEDs.
Measuring interval
(*)
1, 2, 5, 10, 15, 30 s / 1, 2, 5, 10, 15, 30, 60 min
Logging and
transmission interval
(*)
1, 2, 5, 10, 15, 30 s / 1, 2, 5, 10, 15, 30, 60 min
Internal memory
Circular management or stop logging if memory is full.
The number of storable samples depends on the number of detected
quantities (see table 8.7.1).
Alarm
Acoustic by means of the internal buzzer
Power supply
Non rechargeable lithium thyonil chloride (Li-SOCl
2
) internal battery, 3.6
V, size A, 2-pole Molex 5264 connector.
Battery life
(without repeaters,
direct communication
with HD35AP...)
1.5 years typical for CO/CO
2
models (with 2 min measurement and logging
intervals) and for
∆
P range r5 model (with 30 s measurement and logging in-
tervals);
2 years typical for the other models, with logging interval 30 s and measure-
ment interval 30 s for
∆
P ranges r1…r4 models, 10 s for HD35EDH and 5 s
for the other models.
Operating temperature
and humidity
-20…+70 °C (-10…+70 °C for models with grid)
0…85
%RH non condensing
Dimensions
See dimensional drawings
Connectors for external
probes with cable
According to the model, M12 connectors or 3.5 mm pitch input terminals.
Weight
200 g approx. (version with LCD, battery included)
Housing
Plastic material
Protection Degree
IP 50 (except versions with grid)
Installation
Wall support (supplied) for removable installation or flanges (option) for
fixed installation.
(*)
Some models that measure several quantities may have a minimum interval exceeding 1 second (see
table 8.7.1).