 Installation and Operating Manual

IP227, Rev DA

Mobrey MCU200 Series 

February 2020

General safety precautions

The equipment described in this manual has been designed in accordance with EN61010 “Safety requirements for 
electrical equipment for measurement, control and laboratory use”, and has been supplied in a safe condition.
To avoid injury to an operator or service technician, the safety precautions given below, and throughout the 
manual, must be strictly followed whenever the equipment is operated, services or repaired.
For specific safety details, please refer to the relevant sections within the manual.The equipment is designed solely 
for electronic measurement and should be used for no other purpose. Delta Mobrey accepts no responsibility for 
accidents or damage resulting from failure to comply with these precautions.

Grounding

To minimise the hazard of electrical shock, it is essential that the equipment be connected to a protective ground 
whenever the power supply, measurement or control circuits are connected, even if the equipment is switched off. 
The electronics unit must be connected to ground using the marked case stud before control or signal leads are 
connected. The ground connections must have a current rating of 25 A.

AC supply

Never operate the equipment from a line voltage or frequency in excess of that specified. Otherwise, the insulation 
of internal components may break down and cause excessive leakage currents.To allow the electronics unit to be 
isolated from the ac supply, the supply must be routed through a switch (or circuit breaker). The switch (or circuit 
breaker) must be within easy reach of the operator and must be clearly identified as the means of supply isolation. 
The maximum current drawn from the supply must be limited by a fuse or trip, to a maximum of 13 A.

Fuses

Before switching on the equipment, check that the fuses accessible from the interior of the equipment are of the 
correct rating. The rating of the ac line fuse must be in accordance with the voltage of the ac supply. If any fuse 
continually blows, do not insert a fuse of a higher rating. Switch the equipment off, clearly label it “unserviceable” 
and inform a service technician.

Explosive atmospheres

Never operate the equipment, or any sensors connected to the equipment, in a potentially explosive atmosphere.
It is not intrinsically safe and could possibly cause an explosion.

Safety symbols

For the guidance and protection of the user, the following safety symbols appear on the equipment:

Symbol:

Meaning:

Fault indicator. Refer to the user manual for detailed instructions of use.

Hazardous voltages

Содержание Mobrey 402SD

Страница 1: ...anual IP227 Rev DA Mobrey MCU200 Series February 2020 Mobrey Ultrasonic Liquid Level Detection Systems for Interface Applications User Manual Control unit types MCU201 and MCU203 Ultrasonic gap sensor...

Страница 2: ...ctions must have a current rating of 25 A AC supply Never operate the equipment from a line voltage or frequency in excess of that specified Otherwise the insulation of internal components may break d...

Страница 3: ...is connected to its supply the opening of covers or removal of parts could expose live conductors The equipment must be disconnected from all power and signal sources before it is opened for any adjus...

Страница 4: ...MCU200 controller B Mobrey 402SD sensor C Mobrey 442SD sensor 2 Sensor installation 2 1 Safety matters Review all the safety information at the front of this manual 2 2 How ultrasonic gap sensors wor...

Страница 5: ...ther side of the gap In this orientation the switching level is half way up the sensor face If the sensor is mounted from the side of the tank the switching level is normally on the centreline of the...

Страница 6: ...cable runs separated by 0 15 metres to minimise cross talk If several sensor cables are being run together then all the transmit cables those connected to E2 should be grouped together and all receive...

Страница 7: ...ified A suitable switch or circuit breaker must be included in the installation and this should be in close proximity to the equipment and marked as its disconnecting device A suitable fuse rated at 3...

Страница 8: ...be used in all applications DPCO Double Pole Change Over relay connections There are two sets of contacts If the relay is connected to hazardous live mains circuits External circuits e g signal circu...

Страница 9: ...de switch is labelled FREQ and is located between the sensor terminal block E 2 and the Aux Input terminals This switch selects the operating frequency of the MCU200 oscillator which is either 3 7 MHz...

Страница 10: ...tory position is to the right for all switches OSC sensor oscillating E relay energised NE de energised Note Slide switches are adjustable using a pencil Ex factory settings are with all switches to t...

Страница 11: ...tch number 4 to the right E GREEN Figure 6 Example time delays for state changes A 2 seconds B 8 seconds C 30 seconds 4 Applications 4 1 Gain adjustment Correct adjustment of the gain is essential for...

Страница 12: ...nsor is normally suspended in the settling tank Similarly the control unit can be used with a Mobrey sludge pipe and a pair of 488S sensors to control the end of a de sludge cycle when thin solids are...

Страница 13: ...ercentage Settled sludge discharge control Reduce the timedelaysetting tomakeadjustments easier See RelayoutputandLEDlogic selection on page 10 for details Set the operating frequency on the control u...

Страница 14: ...Viscous liquids emulsions and liquids containing solid particles have a greater ultrasonic attenuation than clear liquids This technique is used to detect which liquid is present at the sensor e g th...

Страница 15: ...t is split at the interface into reflected and a refracted beams so that it does not reach the receiver crystal If there is no interference in the gap but only one liquid the beam is received and the...

Страница 16: ...can be then checked by linking the two terminals both labelled as E on the sensor terminals This should stop the Fault LED illuminating If the sensor is giving incorrect indications check the gain se...

Страница 17: ...by linking the sensor terminals 1 and 2 with a wire to simulate an oscillating sensor 7 Specifications 7 1 MCU201 Power supply 110 120 10 220 240 10 50 60 Hz Installation categories II IEC60664 for 23...

Страница 18: ...ensors are supported A switch on the PCB selects the electronics to operate at 1 MHz or 3 7 MHz Cable check option is supported for some sensors External input Available for keeping the relay de energ...

Страница 19: ...19 Installation and Operating Manual IP277 Rev DA Mobrey MCU200 Series February 2020...

Страница 20: ...y Limted Riverside Business Park Dogflud Way Farnham Surrey GU9 7SS 44 0 1252 729140 44 0 1252 729168 mobrey delta mobrey com Terms and Conditions of Sale are available upon request All other marks ar...

Отзывы: