![Del Medical EPEX Omniflex IV Скачать руководство пользователя страница 159](http://html1.mh-extra.com/html/del-medical/epex-omniflex-iv/epex-omniflex-iv_service-manual_2465595159.webp)
EPEX/Omniflex IV System Service Manual
Diagnostics and Troubleshooting
7-13
B0
Sync Pulse Error
Press
RESET
(located in the radiography
control group) and try the exposure again.
B1
No Digital Input
• Exposure request timed out
• Bucky contacts no closing
• No exposure taken
1. Exposure never started.
2. Exposure not taken during 8-second
time available.
3. Time out flag.
B2
Abort Exposure
No mAs
Handswitch let off too soon.
B3
X-ray Switch Jammed
Check the handswitch or wiring.
B5
mAs Not Reached and Time
was Stopped
• Too low mA
• Filament out of calibration
• Time exceeded 130% of the
expected mAs
1. Press
RESET
(located in the
radiography control group) and try the
exposure again.
2. Suspect miscalibration.
2B
Portrait/Landscape
Position Fault
Verify Collimator is On.
Table 7-2.
System-Level Error Codes
Error
Error Type
Possible Cause
Recommended Action
Содержание EPEX Omniflex IV
Страница 10: ......
Страница 37: ...EPEX Omniflex IV System Service Manual General Information 2 5 Figure 2 1 Overhead Tube Crane Components...
Страница 84: ...EPEX Omniflex IV System Service Manual 3 36 Installing the System Hardware...
Страница 138: ...EPEX Omniflex IV System Service Manual 4 54 Adjustments and Calibration...
Страница 146: ......
Страница 182: ......
Страница 184: ......
Страница 186: ......
Страница 188: ......
Страница 190: ......
Страница 192: ......
Страница 194: ......
Страница 196: ......
Страница 198: ......
Страница 200: ......
Страница 202: ......
Страница 204: ......
Страница 206: ......
Страница 208: ......
Страница 210: ......
Страница 212: ......
Страница 223: ...EPEX Omniflex System Service Manual Field Replaceable Units D 11...
Страница 224: ......
Страница 226: ......