Error and Warning Reporting
Errors in the DirectRay Controller are reported over the SCSI system
control interface. Errors reported over the system control interface are
classified into two types:
•
Command Errors
–are errors related to commands sent over the
system control interface. The system control interface returns
explanatory information. The error is reported once, then cleared.
Command errors include, but are not limited to:
•
Command field errors
•
Unsupported commands
•
Out of sequence commands
•
Communication protocol errors
•
System Errors
–are due to errors detected in the DirectRay Controller.
In this case, the error status is reported through the system control
interface. System errors prevent “Normal” type scan commands from
being executed. An attempt to send these commands will fail.
Commands that are not related to generating images for medical
diagnostic purposes will work normally.
Содержание EPEX Omniflex IV
Страница 10: ......
Страница 37: ...EPEX Omniflex IV System Service Manual General Information 2 5 Figure 2 1 Overhead Tube Crane Components...
Страница 84: ...EPEX Omniflex IV System Service Manual 3 36 Installing the System Hardware...
Страница 138: ...EPEX Omniflex IV System Service Manual 4 54 Adjustments and Calibration...
Страница 146: ......
Страница 182: ......
Страница 184: ......
Страница 186: ......
Страница 188: ......
Страница 190: ......
Страница 192: ......
Страница 194: ......
Страница 196: ......
Страница 198: ......
Страница 200: ......
Страница 202: ......
Страница 204: ......
Страница 206: ......
Страница 208: ......
Страница 210: ......
Страница 212: ......
Страница 223: ...EPEX Omniflex System Service Manual Field Replaceable Units D 11...
Страница 224: ......
Страница 226: ......