EPEX/Omniflex IV System Service Manual
Adjustments and Calibration
4-29
Verifying VSID X-ray Field Centered to the DirectRay Detector Center
To verify VSID X-ray field centered to DirectRay Detector center:
Step
Action
1
Using the measure function in DRIVEL, measure from the central
beam spot on the beam alignment tool to all four edges of the black
X-ray field.
Note: The central beam should be centered so that the distances to
all four sides are equal. It is not necessary at this time to align the
X-ray field to the Bucky.
2
Initiate exposure.
3
Accept the image in DRIVEL.
4
Interactively adjust the location of the X-ray field until it is centered
in portrait and landscape.
Note: 21CFR allows a 2% error in this alignment. In order to ensure
image quality and correct alignment, it is suggested that a 1% error
or better (1 cm total at 100 cm) be achieved.
Содержание EPEX Omniflex IV
Страница 10: ......
Страница 37: ...EPEX Omniflex IV System Service Manual General Information 2 5 Figure 2 1 Overhead Tube Crane Components...
Страница 84: ...EPEX Omniflex IV System Service Manual 3 36 Installing the System Hardware...
Страница 138: ...EPEX Omniflex IV System Service Manual 4 54 Adjustments and Calibration...
Страница 146: ......
Страница 182: ......
Страница 184: ......
Страница 186: ......
Страница 188: ......
Страница 190: ......
Страница 192: ......
Страница 194: ......
Страница 196: ......
Страница 198: ......
Страница 200: ......
Страница 202: ......
Страница 204: ......
Страница 206: ......
Страница 208: ......
Страница 210: ......
Страница 212: ......
Страница 223: ...EPEX Omniflex System Service Manual Field Replaceable Units D 11...
Страница 224: ......
Страница 226: ......