EPEX/Omniflex IV System Service Manual
Adjustments and Calibration
4-33
Aligning the HSID Central Beam and X-ray Field Center to DirectRay
Detector Center
To align the HSID central beam:
Step
Action
1
Position the X-ray tube and Bucky to the maximum HISD (A2, B2,
or C2).
2
Align the light field to the test tool.
3
Verify the bead overlap in DRIVEL.
4
If necessary, adjust/compromise positions:
A: Transverse and vertical
B: Longitudinal and vertical
C: Telescope rotation and vertical (transverse only if absolutely
necessary)
5
Verify the bead overlap at HSID Min (A1, B1, or C1) and Max.
6
If necessary, adjust A/B/C: Transverse and telescope, vertical and
tube angle (last).
7
Set detents and reverify VSID Min and Max.
Содержание EPEX Omniflex IV
Страница 10: ......
Страница 37: ...EPEX Omniflex IV System Service Manual General Information 2 5 Figure 2 1 Overhead Tube Crane Components...
Страница 84: ...EPEX Omniflex IV System Service Manual 3 36 Installing the System Hardware...
Страница 138: ...EPEX Omniflex IV System Service Manual 4 54 Adjustments and Calibration...
Страница 146: ......
Страница 182: ......
Страница 184: ......
Страница 186: ......
Страница 188: ......
Страница 190: ......
Страница 192: ......
Страница 194: ......
Страница 196: ......
Страница 198: ......
Страница 200: ......
Страница 202: ......
Страница 204: ......
Страница 206: ......
Страница 208: ......
Страница 210: ......
Страница 212: ......
Страница 223: ...EPEX Omniflex System Service Manual Field Replaceable Units D 11...
Страница 224: ......
Страница 226: ......