Rev 2022-03-08
Pharmafill Model TC4 User Manual
Deitz Company Inc.
TC4 User Manual
Page
33
9.2.
BUILDUP OF REMOVE-TO-CLEAN CONTACT PARTS
Use the pictures in section 9.1.1 (Step-by-step Breakdown…) in reverse order to help with this
procedure.
1.
Install the hopper door (if applicable) – put the door in place and secure with two knobs.
2.
Install the feeder sieve plate (if applicable) – Install sieve plate and tighten two knobs.
3.
Install the funnel block – raise the block as you tighten two knobs.
4.
Install the left, right and center head blocks – slide the blocks onto the mounting posts.
5.
Install the funnel holder and funnel – slide in the funnel holder and tighten the knob.
6.
Install the counting head clear front cover – install the cover and tighten the three knobs.
7.
Install the counting window and sensor – Install the window and sensor as one piece and
tighten the knob on the left side of the counting window. Install the counting sensor by
snapping it down onto the mounting bracket. Connect the electrical cable by aligning the
notch in the connector to the rear and pushing up onto the sensor. Push the black ring up
to lock the connector.
8.
Install the glass disc – carefully place the glass disc onto the hub, aligning the groove in
the disc support ring with the pin on the hub. Install the cushioning washer and screw
down the hold-down ring nut.
9.
Install the rim band – starting at the front, place the band in the groove, working your
way around to the back. Make sure it is seated down all around.
10.
Install the product guide assembly – position assembly and tighten the knob.
11.
Install the center hub – Place the hub onto the center port, making sure to align the pin in
the hub with the hole in the post. Tighten the knob.
12.
Rotate the air manifold assembly into place– rotate the assembly down and press the
release pin to assure that it has engaged the locating hole.
13.
Install the clear rim cover – Slide the cover onto the rear knobs and tighten. Close the
cover and tighten the front hold-down knob.
14.
If equipped with the optional automatic feeder control, connect the electrical cable to the
AFC sensor by aligning the connector and screwing it on.
Содержание PHARMAFILL TC4
Страница 1: ...Model TC4 Electronic Counter Operation Manual...
Страница 52: ......
Страница 53: ......
Страница 60: ......
Страница 61: ......
Страница 62: ......
Страница 63: ......
Страница 64: ......
Страница 65: ......
Страница 66: ......
Страница 67: ......
Страница 68: ......
Страница 69: ......