Rooster™
Sensor200 USER MANUAL
Rooster™
Sensor200 USER MANUAL 62360MN000-A00
28 of 40
Feature
Object
Type
Device
Instance
Units
Range
Read/
Write
Definition
Default
Value
Notes
NSB High
Alarm Clear
Analog
Value
10
m/s
0 to 40.99
R/W
High air velocity value that will
clear an alarm
No value set Valid in Night Setback mode of
operation
Backlight Dim
Pct
Analog
Value
11
%
5 to 100
R/W
Touch screen dim level as a
percentage of highest brightness
level
50
Backlight
Normal Pct
Analog
Value
12
%
5 to 100
R
Touch screen brightness level as
a percentage
100
Duct Height
Analog
Value
13
cm
0 to 122
R/W
Airflow duct area height
0
Valid if the selected duct area is
rectangle
Duct Width
Analog
Value
14
cm
0 to 183
R/W
Airflow duct area width
0
Valid if the selected duct area is
rectangle
Duct Dia
Analog
Value
15
cm
0 to 183
R/W
Airflow duct area diameter
0
Valid if the selected duct area is
circle
Face Velocity
Analog
Input
0
m/s
0 to 5
R
Real time face velocity with sash
in any position
No value set
Air
Temperature
Analog
Input
1
C
0- 150
R
Air temperature
F
Airflow
Changes Per
Hour
Analog
Input
2
NA
-
R
Number of times the air in a
system has been completely
refreshed in an hour
No value set Valid if Airflow Changes/Hr is
enabled and room volume is
entered
Sash Close
Voltage
Analog
Input
3
V
0
–
10 V
R
Calibrated sash voltage with the
sash closed
Valid if Sash Type is analog
Sash Open
Voltage
Analog
Input
4
V
0
–
10 V
R
Calibrated sash voltage with the
sash in the high position
Valid if Sash Type is analog
Sash Normal
Voltage
Analog
Input
5
V
0
–
10 V
R
Calibrated sash voltage with the
sash in the normal operating
position
Valid if Sash Type is analog