Startup Delay
Definition
30s
Alarm is paused for 30 seconds on power up. This is the default setting.
1 min
Alarm is paused for 1 minute on power up.
2 min
Alarm is paused for 2 minutes on power up.
5 min
Alarm is paused for 5 minutes on power up.
Splash Screen
During startup, the Sensor200 displays the splash screen first and then the home screen. The
user has the ability to display the splash screen throughout the entire startup delay period by
changing Splash Screen from “Normal” to “Startup Delay”.
Home>Settings>System>Alarms/Sound>Alarm Controls>Latch Alarm
Latch Alarm
Alarm latching indicates that an alarm state was triggered even if the device is no longer in an
active alarm state. This could occur when a user has stepped away from the workstation. When
latching is
“On”, a yellow banne
r will display beneath the velocity and temperature readings on
the home screen to indicate that a low airflow event has occurred. When latching is set to
“Password”, an EH&S passcode must be entered to unlatch this alarm. If the Latch alarm setting
is set
to “On”, tap the banner to disable the alarm
. Setting Alarm Latch to
“Off” disables this
feature entirely.
Home>Settings>System>Alarms/Sound>Alarm Controls>Latch Alarm
Alarm Tone Settings
The user can configure alarms with different tones. Alarm tone settings are available for Low
Flow, Sensor, and Sash Alarms. At the time of tone selection, the tone plays for approximately 2
seconds to understand the tone type and volume level. Default alarm tone settings are:
▪
Low Flow alarm - Tone 1
▪
Sensor alarm - Off
▪
Sash alarm - Tone 2
Alarm Tone settings supersede Minimum Volume settings.
Alarm Type
Definition
Off
The audible alarm is off.
Tone 1
Alarm plays Tone 1.
Tone 2
Alarm plays Tone 2.