19
KD2 Pro Operator’s Manual
3. The Menus
Repeat this until you reach the last character. When finished,
simply continue to press Enter until you reach the end of the
row. Holding Enter or Escape will allow you to scroll left and
right through the Sample ID quickly.
NOTE: It is important to wait for about fifteen minutes
between readings, if the sensor is left in the same loca-
tion!
If multiple measurements of a sample are made too rap-
idly in succession, the sample’s temperature will not have had
enough time to equilibrate from the previous reading, and the
resulting measurement will be inaccurate. For best results, the
sample should be as close to equilibrium as possible. A good
environment for equilibration can be provided by placing the
KD2 Pro sensor and sample in an isothermal chamber or sty-
rofoam box.
The Data Menu
The Data Menu allows you to view stored data, download data
to a computer, and erase the data from your KD2 Pro. It also
displays how many readings have been recorded out of the
total 4,095 that can be stored.
Here is description of each sub-menu:
VIEW:
This menu allows you to view all data currently stored
on your KD2 Pro. Press Enter to access the readings, and use