10
KD2 Pro Operator’s Manual
2. KD2 Pro Overview
KD2 Pro Sensor Compatibility
1
In low viscosity liquids, the read time should be set to the
minimum allowed 1 minute and should only be used in low
power mode to avoid free convection.
2
In high viscosity liquids, the KS-1 or TR-1 sensor should
only be used in low power mode.
3
The KS-1 or TR-1 needles in low power mode with 10 min
read time give good results with insulation. Heater tempera-
tures with high power mode are quite high for insulation mea-
surements.
Sample Material
KS-1
TR-1
SH-1
Low viscosityliquids
(e.g. water)
Best
1
No
No
High viscosity liquids
(e.g. glycerol, oil)
Best
2
Ok
2
No
Insulation and insulating
materials
Best
3
Ok
Ok
Moist soil
No
Best
Ok
Dry soil, powders, granu-
lar materials
No
Best
4
Ok
Concrete
No
Best
5
Ok
6
Rock
No
Best
5
Ok
6
Other solids
No
Best
5
Ok
Volumetric specific heat
No
No
Best
Thermal diffusivity
No
No
Best