General Service Guide
73
Document No. 896 623
79.
Check the communication to the network by checking the states of the Network Board rear
panel LEDs:
TxD (Green); should blink intermittently
RxD (Yellow); should blink intermittently
L (Green); should be lit continuously
C (Yellow); should be lit continuously
NOTE: The B-UPINET board doesn’t have LEDs.
?
80.
Enter the
Network
service menu:
Monitor Setup
- Install/Service
(password 16-4-34)
-
Service
(password 26-23-8)
- Frame - Network
Check that the counters for data errors (‘CRC’, ‘Frame’, ‘Transm.’) are stable.
NOTE: The counters may show values greater than 0. However, any values increasing
continuously indicates a problem.
?
81.
Check that the counters for hardware errors (‘Intern.’, ‘Missed’, ‘FIFO’, ‘Overrun’) all show 0.
If any of the counters show a value greater than 0, replace the Network Board.
?
82.
Turn the monitor off and perform final cleaning.
?
•
Fill in all necessary documents.