General Service Guide
29
Document No. 896 623
3.6
Network Board, B-NET Installation (w/ B-CPU2/3)
Insert the Network Board, B-NET according to following procedure.
1.
Make sure that the power is turned off on the Compact Monitor is and unplug the Power
Cord. Press and hold the Service Reset button on the rear panel for at least five seconds
until the service reset indicator LED is switched off.
2.
Unscrew the six screws holding the back plate in position and remove the plate.
WARNING
The Compact Monitor is always energized by the internal battery. Do not touch
any exposured wiring or conductive surface with metallic object while the back
plate is off or electric failure may result.
CAUTION
The Compact Monitor is always energized by the internal battery. A short circuit may
cause internal damage.