Datex-Ohmeda AS/3 Compact Monitor, CS/3 Compact Monitor
22
Document No. 896 623
3.
Remove the software cartridge.
4.
Unscrew the screw holding the CPU Board in position and remove the board.
5.
Remove the CPU Board from the protective anti static packaging. Always hold the board by
the edges and wear a wrist grounding strap.
CAUTION
The CPU Board comprises sensitive integrated circuits that can be damaged by an
electrostatic discharge. Careful handling of the board is therefore essential.
6.
Remove the EMC plate from the CPU Board.
7.
Insert the new CPU Board into the empty slot and firmly press the board in position. Secure
the new board.
8.
Replace the Software Cartridge and secure the cartridge with the two screws removed from
the cartridge earlier.
9.
Replace the back plate and secure the plate with the six screws removed from the plate
earlier.
10.
Reconnect the Power Cord.
11.
Perform Factory Reset.