Aestiva
A-18
1006-0938-000
Minimize alarms
(optional)
Set the ventilator controls to decrease the number of alarms:
1.
Control Keys:
•Volume alarms: Off
•Plimit: 100 cmH
2
O
2.
Alarm menu:
•Low O
2
: 21%
•High O
2
: Off
3.
Bag/Vent switch: Bag
After you have finished testing, set the Plimit back to a clinically appropriate
value.
Vaporizer installation
w
W
W
W
WA
A
A
AR
R
R
RN
N
N
NIIIIN
N
N
NG
G
G
G
Use only the Tec 5, Tec 6, and Tec 7 vaporizers.
Do not use a vaporizer that lifts off the manifold when the lock lever
is in the locked position.
Do not use this anesthesia system if you can turn on more than one
vaporizer at the same time.
Tec 6 vaporizers will not align correctly unless the power cable goes
through the channel on the bottom of the vaporizer. Do not put the
power cable on top of the manifold or between vaporizers.
The vaporizer storage bracket is not part of the vaporizer manifold.
You cannot use a vaporizer while it is on the bracket.
1.
If the top of a vaporizer is not horizontal, remove the vaporizer and
reinstall it.
2.
Set each vaporizer lock lever to the locked position.
3.
Try to lift each vaporizer off of the manifold.
4.
If a vaporizer lifts off the manifold, install it again and complete steps 1,
2, and 3. If the vaporizer lifts off a second time, do not use the system.
Содержание 2 Vap. Trolley
Страница 1: ...Aestiva 5 Operation Manual Part 1 Software Revision 4 X System Controls Operation Checkout ...
Страница 6: ...Aestiva iv 1006 0938 000 iv ...
Страница 14: ...Aestiva 1 8 1006 0938 000 1 8 ...
Страница 36: ...Aestiva 2 22 1006 0938 000 ...
Страница 76: ...Aestiva 4 4 1006 0938 000 ...
Страница 102: ...Aestiva I 4 1006 0938 000 ...