3 Operation and Tutorial
1006-0938-000
3-11
3
Set an audible alarm for circuit leaks
The patient circuit leak alarm is activated if less than half of the inspired volume
returns through the expiratory flow sensor during mechanical ventilation. To
prevent nuisance alarms from a known leak (e.g., an un-cuffed endotracheal
tube), set the audio to Off. Normal volume and apnea monitoring does not
change.
N
N
N
Noo
o
ottttee
e
e::::
This alarm is the first stage in detecting a circuit disconnect. The audible
leak alarm cannot be turned off unless volume alarms are on, and the low
◊
E
limit is set to a value other than off.
Step 1
Select the Alarm Settings menu.
• Push the menu key.
• Turn, then push the knob to select
Alarm Settings.
AB.29.013
AB.90.036
AB.90.038