Aestiva
3-12
1006-0938-000
Set Cardiac Bypass
Set Cardiac Bypass to In Progress to prevent volume and apnea alarms when
the patient is on cardio-pulmonary bypass.
When Cardiac Bypass In Progress is selected, the display shows:
• Cardiac Bypass
• Apnea Alarm Off
• Vol Alarms Off
Note
::::
The mechanical ventilation must be off. When the mechanical
ventilation is turned back on, Cardiac Bypass returns to the No Bypass setting
and alarms become active.
Step 2
Turn, then push the knob to select
circuit leak audio.
Step 3
Turn, then push the knob to change the
setting. You must push the knob to save
the change.
AB.90.054
AB.90.056
Then
Audio On
Audio Off