Appendix - Preoperative Tests
1006-0938-000
A-11
•Make sure that a low minute volume alarm occurs.
•Go to the alarms menu.
•Set the low minute volume alarm to Off.
9.
Test the high airway pressure alarm:
•Set Plimit to less than the peak airway pressure.
•Make sure that the high airway pressure alarm occurs.
•Set Plimit to the correct level.
10. Test the apnea and low airway pressure alarms:
•Remove the test lung from the patient connection.
•Other alarms such as low minute volume can occur.
•Make sure that the low airway pressure and apnea alarms occur. The
apnea alarm occurs after 30 sec.
11. Test the sustained airway pressure alarm:
•Set the controls:
•Mechanical ventilation stops when the Bag/Vent switch is set to Bag.
•Close the patient connection and push the O
2
Flush button.
•Make sure that the sustained pressure alarm occurs after approximately
15 seconds at the sustained pressure limit (6-30 cmH
2
O varies with
pressure limit).
Precase steps
1.
Before the first case, do a low-pressure leak test. Refer to “Every time a
different clinician uses the system.”
2.
Do the last two sections of tests in “Before every patient”:
•Breathing system tests
•Monitor and ventilator tests
APL Valve
Closed
Bag/Vent switch:
Bag