3 Operation and Tutorial
1006-0938-000
3-31
3
Show the service settings
The About Ventilator screen shows ventilator settings that can only be changed
by an approved service representative.
Item
Use
Software Version
If you call for service, a representative may ask for
this.
Facility defaults or last
settings
Tells you if the system saves the current settings
when you turn it off or goes back to facility defaults.
Altitude
Used for gas calculations. If the altitude is not
correct, O
2
calibration can fail.
Drive gas (O
2
or Air)
Tells you which gas the ventilator uses to drive the
bellows. This gas comes from the same supply
(pipeline or cylinder) that the anesthesia machine
uses. If this gas comes from a cylinder, the cylinder
empties faster than you would expect from the
flowmeter settings.
Step 1
Select the Setup/Calibration
menu.
• Push the menu key.
• Turn, then push the knob to
select Setup/Calibration.
AB.29.013
AB.90.036
AB.90.037