2 System Controls and Menus
1006-0938-000
2-21
2
Optional CO
2
Bypass mode operation
Pulling and turning the canister release opens the canisters and activates the
optional CO2 bypass mode. The CO2 bypass seals the breathing circuit when
the canisters are open. This permits continued ventilation and rebreathing of
exhaled gases.
When the canisters are open, the medium priority alarm message “No CO2
Absorption” is displayed. The alarm is downgraded to low priority after silenc-
ing.
Close the absorber canisters to activate the Absorber mode. Closing the can-
isters directs exhaled gas flow through the absorber, removing CO2. Do the
breathing system tests as soon as possible after returning to Absorber mode.
When operating in CO2 Bypass mode, water condensate may collect in the
bypass assembly. The water drains automatically when the canisters are
closed.
Datex-Ohmeda strongly recommends using CO2 monitoring when using the
CO2 Bypass device.
Systems which have the optional CO2 Bypass mode installed have the follow-
ing label at the canister release.
AB.23.
1
7
9