Optional Passive AGSS operation
The Aestiva/5 7100 Passive AGSS consists of
a two-liter reservoir to minimize rapid
pressure changes. It contains both positive
and negative pressure relief valves to protect
the breathing system. The outlet is a 30 mm
male taper swivel connector (a) at the rear of
the breathing system.
The connector also has a capped hose barb that may be used for scavenging
the sample from a gas monitor.
Passive AGSS (Anesthesia Gas Scavenging System) is intended primarily for
use in operating room environments which have no dedicated vacuum system
for waste gas disposal. The disposal system generally consists of large
diameter tubing and/or duct directly linking the passive AGSS with the
building exterior. The tubing should be as large in diameter and as short as
needed for the particular application.
Passive AGSS may also be used with a non-recirculating ventilation system for
waste gas disposal. The tubing connection from passive AGSS to the non-
recirculating ventilation system should be an open connection, essentially at
atmospheric pressure, such as to an exhaust grill.
Passive AGSS may also be used as a protective interface to an externally
mounted active AGSS such as the Datex-Ohmeda adjustable Waste Gas
Scavenging Interface Valve Assembly.
a
AB.23.155
Содержание 2 Vap. Trolley
Страница 1: ...Aestiva 5 Operation Manual Part 1 Software Revision 4 X System Controls Operation Checkout ...
Страница 6: ...Aestiva iv 1006 0938 000 iv ...
Страница 14: ...Aestiva 1 8 1006 0938 000 1 8 ...
Страница 36: ...Aestiva 2 22 1006 0938 000 ...
Страница 76: ...Aestiva 4 4 1006 0938 000 ...
Страница 102: ...Aestiva I 4 1006 0938 000 ...