M-Class
75
Advanced Entry Calibration (continued)
Step
Operator Action
Displayed Message
Comment
12
From the data collected in
Steps 10 and 11, where both
the sensor readings are above
20, subtract each Label Value
from the corresponding TOF
Value. These are the
Difference Values
.
From the resulting list, find the
largest Difference Value (see
example below). Its associated
Gain Number will provide the
best algorithm for your media.
!
HBJO!USBO!!!!!!=zzz?+1
1!!!!!!!!!)1!.!42*!
!
Or, for reflective media:
HBJO!SFGM!!!!!!=zzz?+1
1!!!!!!!!!)1!.!42*!
Both sensor readings must be above 20.
Gain
Number
Label
Value
TOF
Value
Difference
Value
00 255 254 1
01 251 240 11
02 241 213 28
03 231 182 49
04 219 150 69
05 212 119 93
06 200 88 112
07 189 58 131
08
178
32
146
09 167 19
N/A
10 156 17
N/A
11 146 16
N/A
12 136 15
N/A
… … … …
31 116 14
N/A
For example, if your
compiled data had the
values shown in this
table, Gain Number 8
would be chosen
because it has the
highest Difference
Value (146) where
both the Label Value
and the TOF Value
are above 20.
32 112 14
N/A
Содержание M-Class M-4206
Страница 1: ...M Class Operator s Manual ...
Страница 10: ...vi ...
Страница 26: ...16 M Class ...
Страница 48: ...38 M Class ...
Страница 88: ...78 M Class ...
Страница 112: ...102 M Class ...
Страница 124: ...114 M Class Bar Code W1T TCIF Linked Barcode 3 of 9 TLC39 ...
Страница 130: ...120 M Class Network Status ...
Страница 131: ...M Class 121 TCP IP Configuration ...
Страница 133: ...M Class 123 System Settings Media Settings and Print Control continued Printer Options Communications and Diagnostics ...
Страница 134: ...124 M Class Printer Options Communications and Diagnostics continued ...
Страница 135: ...M Class 125 Network Print Options Reset Network Parameters ...
Страница 136: ...126 M Class Change Password ...
Страница 156: ...146 M Class ...
Страница 162: ...152 M Class ...
Страница 168: ...158 M Class ...