TECHNICAL SUPPORT
70
UNIQ™
List of problems related to laser marker states
POSSIBLE CAUSE
ACTION
Blinking
GREEN
- Lighter™ Suite marking SW
corrupted
- C:\ or D:\ drive corrupted
Restore the laser marker (see “Recover the
laser marker” on page 106)
Steady
GREEN
EXT_KEY
contact is LOW level
or disconnected
Check
X1.12
(
EXT_KEY
) input signal on the
Command Box connector is set to
HIGH
level (see “X1 - Command Box (Laser Con-
trol)” on page 25)
Steady
ORANGE
EXT_ENABLE_A
and/or
EXT_ENABLE_B
contact are
LOW level or disconnected
Check
X1.8
(
EXT_ENABLE_A
) and
X1.2
(
EXT_ENABLE_B
) input signals on the
X1 -
Command Box
connector are set to
HIGH
level (see “X1 - Command Box (Laser Con-
trol)” on page 25)
List of most common problems
PROBLEM
DESCRIPTION
STATUS LED
Laser marker never goes to
WAIT FOR START state
Laser marker never goes to
STANDBY SHUTTER CLOSED
state
Laser marker never goes to
READY state
PROBLEM DESCRIPTION
POSSIBLE CAUSE
ACTION
Laser marker doesn't turn ON
Power supply cable disconnected
Check the
POWER SUPPLY CABLE
connec-
tion to the power grid
SWITCH of power supply set to OFF
Check that the
POWER SUPPLY
switch is
set to
ON
FUSES damaged
Check that
FUSES
of power supply are not
damaged
Impossible to control the marking process
using the X1 - Command Box control
signals
Incorrect integration
Check that the
X1 - Command Box
control
signals are compatible with the external
devices used for integration
Bad marking result
F-Theta scan lens is dirty
Clean the
F-Theta scan lens
. See “F-Theta
scan lens cleaning procedure” on page 67
Incorrect focus distance
Check that the material to be marked is
placed at the right working distance. See
“Marking Area Specification” on page 19
Incorrect laser parameters
Check that the laser parameters set in the
layout are appropriate for the material to
be marked. Contact Datalogic Technical
Support
Laser marker doesn't engrave
Incorrect laser marker state
Check the laser marker state is set to
READY
F-Theta lens protection not removed
Check that the F-Theta scan lens protec-
tion has been removed
The optical path is obstructed
Check that the optical path is free
Laser marker doesn't start to engrave
using external START_MARKING signal on
the Command Box connector
STOP_MARKING signal is active
Check that the
X1.13
(
STOP_MARKING
)
signal on the Command Box connector is
not set to
HIGH
level
START_MARKING signal incorrect timing
- Check the
X1.11
(
START_MARKING
) sig-
nal on the
X1 - Command Box connector
(see “X1 - Command Box (Laser Control)”
on page 25).
HIGH
level pulsed signal start
the marking process (refers to Lighter™
user's manual “Setting I/O parameters”
paragraph to set the signal properties)
- Use Lighter™ System Test to check
external
X1.11
(
START_MARKING
) and
X.13
(
STOP_MARKING
) signals
Lighter™ is not set in AUTO MODE /
WORK MODE
Check that Laser Engine is set in
AUTO
MODE
or that Laser Editor is set in
WORK
MODE
Содержание UniQ 1150-1X41
Страница 1: ...UniQ PRODUCT REFERENCE GUIDE InfraRed Fiber Laser Marker ...
Страница 8: ...PREFACE viii UNIQ ...
Страница 82: ...TECHNICAL SUPPORT 74 UNIQ ...
Страница 84: ...LABELS 76 UNIQ POSITIONING OF EXTERNAL LABELS Figure 50 External labels location ...
Страница 120: ...RECOVER THE LASER MARKER 112 UNIQ 16 Press FINISH to continue 17 Press YES to continue ...
Страница 124: ...116 UNIQ APPENDIX G MECHANICAL DRAWINGS ...
Страница 125: ...LASER MARKER PRODUCT REFERENCE GUIDE 117 LASER MARKER NOTE Please refer to Datalogic website for detailed drawings ...
Страница 126: ...MECHANICAL DRAWINGS 118 UNIQ ...
Страница 127: ......