continue from that location.
In this example the circled patch does not fit the pattern of this page, indicating a
double reading in a previous patch. This type of error will cause the end of row tone to
sound at the wrong time, cuing the user that an error has occurred. Use the arrow
keys to return to the beginning of this row, and reread patches until the incorrect area
is overwritten.
When all patches (and for multipage templates, all pages) have been measured a
dialog box appears noting completion of measurements. It is recommended at this
point that the Patch Display mode be changed to Measured mode, and all measured
values be reviewed visually for possible errors. If a single patch does not align
properly with the gradients surrounding it, arrow to that location, and remeasure that
patch to see if remeasurment improves the result. It is possible that the Spectro was
not flush with the surface of the paper, or not entirely on the color patch, causing an
incorrect reading. If remeasurement does not change the value, then the variation is
probably correct, and may be visible in the printed target as well.
In this example, the circled patch is too dark, and does not fit the pattern of the
adjacent patches. It should be remeasured to check that this is not a patch reading
error.
Saving the Measurements