User Guide
MES 1003 WMS
Doc no. 100837-MAN – Rev. E1 – Rev. date 2022-10
© Danfoss IXA A/S | Page 38
Illustration 44: Calibration of pH 7 succeeded
Calibration of pH 7 has succeded. Press OK to continue to calibration with pH 4 buffer solution.
Illustration 45: Clean sensor in distilled water
Remove the pH sensor from the pH 7 buffer solution and clean the sensor in distilled water. Press OK when done.
Illustration 46: Immerse sensor in pH 4 buffer solution
The pH sensor should be immersed in pH 4 buffer solution after the sensor has been cleaned in distilled water. Refer to
to see how to setup the pH sensor and buffer solution for calibration of pH. Press OK when done.
Illustration 47: Calibrating the pH sensor
Wait for the calibration to finish
pH Calibration
6-**
Calibration of pH 7 succeeded
Press OK when done
pH Calibration
6-**
Please clean the sensor in
distilled water
Press OK when done
pH Calibration
6-**
Calibrating pH 4
Immerse the sensor into pH
4 buffer solution
Press OK when done
pH Calibration
6-**
Calibrating…